background image

 

MAINTENANCE 

 

The 

TOBI HOSPI PLUS

 suction equipment does not need maintenance or lubrication. 

It is necessary to check functioning and instrument before every use.                                            
Unpack the instrument and 

always check 

integrity of plastic parts and feeding cable, they might have been 

damaged during previous use. 
Connect cable to electrical network and turn switch on. 
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check that 
the vacuum indicators reaches -90 kPa (-0.90 bar) maximum. 
Rotate the knob from right to left and check the aspiration regulating control.                               
The vacuum indicator should go down -40 kPa (-0.40 bar). 
Verify that loud noises are not present, these can indicate wrong functioning. 
A protection fuse (

F 1 x 4A 250V

) reachable from exterior and it situated in the plug protects the instrument. 

For use replacing, always check the type and the range indicated. 

 

Fault type

 

Cause

 

Solution

 

1. No aspiration 

Jar Cap badly screwed 

down 

Unscrewed the cap, then rescrew it correctly 

2. No aspiration 

Lid seal not in its seat 

Unscrew the cap and insert the seal properly 

in its seat 

3. The float doesn’t close 

If the cap has been washed, 

ensure that the float is not 

partially detached 

Insert the float into it’s place 

4. The float doesn’t close 
 
 

The float it’s covered by 

dirty material 

Unscrewed the cap, leave the and put in on 

autoclave 

5. Low suction 

Foam inside the jar 

Fill the jar to 1/3 full of ordinary water  

6. No aspiration due to   
    flow leakage of mucus  

Filter blocked 

Replace filter 

Faults 1 - 2 - 3 - 4 - 5 - 6 

None of the remedies has 

achieved the desired 

results 

Contact the seller or GIMA S.p.A. After-

sales Assistance Service 

 

If the overfill security system it’s activated, don’t proceed with the liquid aspiration. 
If the overfill security system doesn’t work there are two cases: 
1° case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter 
who avoid the liquid penetration inside the device. 
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device, 
in this case return the device to GIMA S.p.A. technical service. 
 

 

BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, 

PLEASE CONTACT GIMA S.p.A. TECHNICAL SERVICE. 

GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL 

SERVICE CHECKING, APPEARS TO BE TAMPERED. 

 
 
 
 
 
 
 
 
 
 
 
 
 

 12

Summary of Contents for TOBI CLINIC

Page 1: ...MANUALE D USO USER MANUAL MODE D EMPLOI HANDBUCH MANUAL DE ISTRUCCIONES M28205 M Rev 2 10 07 GIMA S p A Via Monza 102 20060 Gessate MI Italia ITALIA Tel 02 953854 1 Fax 02 95381167 e mail gima gimaita...

Page 2: ...rolunghe Qualora il loro utilizzo fosse indispensabile necessario utilizzare tipi conformi alle norme di sicurezza facendo comunque attenzione a non superare i limiti massimi di alimentazione sopporta...

Page 3: ...E TRASPORTO Temperatura ambiente 40 70 C Percentuale umidit ambiente 10 100 RH SIMBOLOGIA ADOTTATA SULL APPARECCHIO Apparecchio con Classe di isolamento II Marchio di conformit alla direttiva 93 42 C...

Page 4: ...Il Cliente e o l utente dell aspiratore chirurgico TOBI HOSPI PLUS devono assicurarsi che l apparecchio sia utilizzato in tale ambiente Test di Emissione Livello di prova Guida all ambiente elettroma...

Page 5: ...vaso parte conica con 3 aperture 3 Dopo aver tolto la gabbietta il galleggiante con guarnizione rossa rimosso da tappo vaso 4 Per disinfettare gli accessori utilizzare alcool denaturato o soluzione a...

Page 6: ...chio del vaso 2 Mancata aspirazione Guarnizione del coperchio non in sede Svitare il coperchio e riposizionare la guarnizione nella sede del coperchio 3 Galleggiante bloccato Incrostazioni sul gallegg...

Page 7: ...ere l apparecchio premere l interruttore sulla posizione 0 ed estrarre la spina dalla presa di alimentazione Estrarre gli accessori e procedere alle operazioni di pulizia E disponibile anche la versio...

Page 8: ...d or extension adapters is not generally recommended Whenever their use is indispensable use those in compliance with safety regulations however paying attention not to exceed the maximum power supply...

Page 9: ...CONDITION Room temperature 10 40 C Room humidity percentage 20 85 RH Altitude 0 2000m s l m CONSERVATION CONDITION AND TRASPORT Room temperature 40 70 C Room humidity percentage 10 95 RH SYMBOLS Clas...

Page 10: ...mers or the user of the surgical aspirator TOBI HOSPI PLUS should assure that it s used in such an environment Immunity Test Compliance Electromagnetic environments guidance Electrostatic discharge ES...

Page 11: ...4 To disinfect the accessories use either denatured alcohol or hypochlorate based solutions easily purchasable from chemist shops 5 Once cleaning has been completed insert the float in the cage with t...

Page 12: ...properly in its seat 3 The float doesn t close If the cap has been washed ensure that the float is not partially detached Insert the float into it s place 4 The float doesn t close The float it s cov...

Page 13: ...tch on O position and unplug To extract the accessories and start with cleaning it The dual use version is available on request Continuous use can be enabled by activating the illuminated switch Alter...

Page 14: ...mplacer la fiche avec une autre d un type ad quat Si leur utilisation est indispensable il faut utiliser des types conformes aux normes de s curit en faisant toutefois attention ne pas d passer les li...

Page 15: ...p rature ambiante 5 35 C Pourcentage humidit ambiante 30 75 RH Altitude 0 2000m s l m CONDITIONS DE CONSERVATION ET DE TRASPORT Temp rature ambiante 40 70 C Pourcentage humidit ambiante 10 100 RH SYMB...

Page 16: ...ou l utilisateur de l aspirateur TOBI HOSPI PLUS doivent s assurer que l appareil est utilise dans un tel environnement Test d Emission Niveau de test Guide l environnement lectromagn tique D scharge...

Page 17: ...vertures 3 Apr s avoir enlev la cage le flotteur avec joint rouge doit tre enlev du couvercle du bocal 4 Pour d sinfecter les accessoires utiliser de l alcool d natur ou une solution base d hypochlori...

Page 18: ...as Joint du couvercle pas en place D visser le couvercle et remettre le joint en place 3 Le flotteur bloqu Pr sence des incrustations sur le flotteur D visser le couvercle et mettre la garniture dan l...

Page 19: ...position 0 et d brancher la fiche de la prise d alimentation Sortir les accessoires et proc der aux op rations de nettoyage Sur demande la version double utilisation est disponible L utilisation cont...

Page 20: ...endete Steckertyp denen des Stromnetzes entsprechen an das man das Ger t anschlie en will 3 Folgende Punkte m ssen besonders beachtet werden Verwendedn Sie nur Originalzubeh r Das Ger t darf nur mit d...

Page 21: ...Betrieb Dauerbetrieb Abmessungen 470 x 740 h x 430mm Betriebsbedingungen Raumtemperatur 5 35 C Raumfeuchtigkeit 30 75 RH H he 0 2000m s l m Lagerung Raumtemperatur 40 70 C Raumfeuchtigkeit 10 100 RH...

Page 22: ...oder Benutzer des Absauger TOBI HOSPI PLUS muss sich vergewissern dass das Ger t unter diesen Bedingungen benutz wird Emissionstests Pr fniveau F her elektromagnetische Umgebung Elektrosatische Entla...

Page 23: ...ch dem Entfernen des K figs wird der Schwimmer mit der roten Dichtung vom Beh lterverschlu entfernt 4 Zum Desinfizieren der Zubeh rteile ein handels bliches mildes Desinfektionsmittel verwenden 5 Nach...

Page 24: ...alls das berlaufschutzystem nicht funktioniert gibt es zwei m gliche Gr nde 1 Wenn das berlaufschutzystem nicht arbeitet wird die Absaugung durch den Bakterienfilter gestoppt der das Eindringen von Fl...

Page 25: ...Zubeh rteile herausnehmen und reinigen TOBI HOSPI PLUS Chirugiesauger PROFI Durch Bet tigung eines Lichtschalters wird der Dauergebrauch erm glicht Alternativ kann der Bediener bei Bet tigung der Ped...

Page 26: ...los datos el ctricos indicados en la etiqueta de los datos y el tipo de enchufe utilizado correspondan a los de la red el ctrica a la cual se conectar 3 En el caso que el enchufe en dotaci n del apara...

Page 27: ...ac on 230V 50Hz Potenzia Absorbida 300 VA Fusible F 1 x 4A 250V Aspiraci n M xima sin vasos 90 kPa 0 90 Bar Flujo M ximo de aspiraci n sin vasos 90 l min Peso 20 Kg Dimensi n 470 x 740 h x 430mm Funci...

Page 28: ...idad Test de Inmunidad Descargas electroest ticas ESD IEC EN 61000 4 2 6kV en contacto 8kV en aire El pavimento deber a ser de madera cemento o cer mica Si el pavimento est recubierto por material sin...

Page 29: ...aperturas 3 Despu s de haber quitado la jaula se quitar el flotante con junta roja del tap n vaso 4 Para desinfectar los accesorios utilizar alcohol desnaturalizado o soluci n a base de hipoclorito 5...

Page 30: ...oscar correctamente 2 Falta aspiraci n Junta del tap n fuera del alojamiento Desenroscar el tap n y acomodar la junta en su alojamiento 3 Falta cierre del flotante Si el tap n ha sido lavado controlar...

Page 31: ...osici n 0 y extraer el enchufe de la toma de alimentaci n Extraer los accesorios y llevar a cabo la limpieza Sobre pedido puede suministrarse en la versi n de double uso Puede as utilizarse en continu...

Page 32: ...DIREZIONE FLUSSO FLOW DIRECTION Montaggio Filtro Filter Assembling Mod TOBI CLINIC SUPER TOBI CARRELLATO Cod 28239 TOBI HOSPITAL TOBI HOSPI PLUS Cod 28237 DIREZIONE FLUSSO FLOW DIRECTION Connessione V...

Reviews: