Immunity test
Test level to
IEC 60601-1-2
Conformity level
Electromagnetic environment -
directives
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic
tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30%
Electrical fast transient
/ burst IEC 61000-4-4
+/- 2 kV
for power supply unit
+/- 1 kV
for input/output lines
+/- 2 kV
for power supply lines
+/- 1 kV
for input/output lines
Mains power quality should be that of a
typical commercial or residential
environment.
Surge
IEC 61000-4-5
+/- 1 kV
differential mode
+/- 2 kV
common mode
+/- 1 kV
differential mode
+/- 2 kV
common mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage
dips,
short
interruptions
and
voltage variations on
power
supply
input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
For 0,5 cycle
40% of U
T
(60% dip in U
T
)
For 5 cycles
70% of U
T
(30% dip in U
T
)
For 25 cycles
<5% U
T
(>95% dip in U
T
)
For 5 sec
<5% U
T
(>95% dip in U
T
)
For 0,5 cycle
40% of U
T
(60% dip in U
T
)
For 5 cycles
70% of U
T
(30% dip in U
T
)
For 25 cycles
<5% U
T
(>95% dip in U
T
)
For 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Product requires continued
operation during power mains interruptions, it
is recommended that the Product be powered
from an uninterruptible power supply or
battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE U
T
is the a.c mains voltage prior to application of the test level.
Summary of Contents for PRIMALED SO
Page 15: ...Operator s Manual MO080 IT 12 09 2017 Rev 3 Page 15 of 20 8 Wiring diagram ...
Page 21: ...Drawing code Rev Data 052 1 2 3 ...
Page 22: ...Drawing code Rev Data 053 1 2 3 ...
Page 23: ...Drawing code Rev Data 054 2 1 3 4 ...
Page 25: ...Drawing code Rev Data 056 2 1 3 4 5 ...
Page 26: ...Drawing code Rev Data 057 3 1 2 ...
Page 27: ...Drawing code Rev Data 058 2 1 3 4 ...
Page 28: ...Drawing code Rev Data 1 2 7 3 4 5 6 153 ...
Page 29: ...Drawing code Rev Data 190 1 2 3 4 5 6 8 7 ...
Page 30: ...Drawing code Rev Data 191 1 2 4 5 3 6 7 ...
Page 31: ......
Page 32: ......