7. Troubleshooting
Trouble
Possible Reason
Solution
Display direction
doesn’t change
or changes
insensitively.
The SpO
2
and Pulse Rate
display unstable
Can not turn on
the device
No display
No display of the
wireless icon “
”
Maybe the oximeter
is not used for a long
time, the movable metal
ball within the
orientation-sensor
can not move freely.
1. The finger is not placed
far enough inside.
2. The finger is shaking
or the patient
is moving.
1. The batteries are
drained or almost
drained.
2. The batteries are not
inserted properly.
3. The device
is malfunctioning.
1. The device will power
off automatically when
it gets no signal for
8 seconds.
2. The batteries
are almost drained.
Hardware failure of
wireless transmission
function.
Please shake the
oximeter with a certain
force to make the
movable metal ball
move freely.
If the problem still
exists, maybe
the orientation-sensor
is not working properly.
Please contact the
local service center.
1. Place the finger
correctly inside
and try again.
2. Let the patient
keep calm.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local
service center.
1. Normal.
2. Change batteries.
Please contact the local
service center.
Summary of Contents for OXY-10
Page 145: ...145 ARABIC ...
Page 146: ...146 ARABIC ...
Page 147: ...147 ARABIC ...
Page 149: ...149 ARABIC ...
Page 150: ...150 ARABIC ...
Page 151: ...151 ARABIC ...
Page 152: ...152 ARABIC ...
Page 153: ...153 ARABIC ...
Page 154: ...154 ARABIC ...
Page 155: ...155 ARABIC ...
Page 156: ...156 ARABIC ...
Page 157: ...157 ARABIC ...
Page 158: ...158 ARABIC ...
Page 159: ...159 ARABIC ...
Page 160: ...160 ARABIC ...
Page 161: ...161 ARABIC ...
Page 162: ...162 ARABIC z ...
Page 163: ...163 ...
Page 164: ......