hemoglobin (HbO
2
) and deoxygenated hemoglobin (Hb) have different
absorption character in the spectrum range from red to infrared light
(600nm~1000nm wavelength), by using these characteristics, SpO
2
can be
determined. SpO
2
measured by this oximeter is the functional oxygen
saturation - a percentage of the hemoglobin that can transport oxygen.
In contrast, hemoximeters report fractional oxygen saturation - a percentage
of all measured hemoglobin, including dysfunctional hemoglobin, such
as carboxyhemoglobin or metahemoglobin.
Clinical application of pulse oximeters: SpO
2
is an important physiological
parameter to reflect the respiration and ventilation function, so SpO
2
monitoring used in treatment has become more popular. (For example,
such as monitoring patients with serious respiratory disease, patients under
anesthesia during operation and premature and neonatal infants) The status
of SpO
2
can be determined in timely manner by measurement and will allow
finding the hypoxemia patient earlier, thereby preventing or reducing
accidental death caused by hypoxia effectively.
3. Factors affecting SpO
2
measuring accuracy (interference reason)
• Intravascular dyes such as indocyanine green or methylene blue
• Exposure to excessive illumination, such as surgical lamps, bilirubin lamps,
fluorescent lights, infrared heating lamps, or direct sunlight.
• Vascular dyes or external used color-up product such as nail enamel
or color skin care
• Excessive patient movement
• Placement of a sensor on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line
• Exposure to the chamber with High pressure oxygen
• There is an arterial occlusion proximal to the sensor
• Blood vessel contraction caused by peripheral vessel hyperkinesias
or body temperature decreasing
4. Factors causing low SpO
2
Measuring value (pathology reason)
• Hypoxemia disease, functional lack of HbO
2
• Pigmentation or abnormal oxyhemoglobin level
• Abnormal oxyhemoglobin variation
• Methemoglobin disease
• Sulfhemoglobinemia or arterial occlusion exists near sensor
• Obvious venous pulsations
• Peripheral arterial pulsation becomes weak
• Peripheral blood supply is not enough
Summary of Contents for OXY-10
Page 145: ...145 ARABIC ...
Page 146: ...146 ARABIC ...
Page 147: ...147 ARABIC ...
Page 149: ...149 ARABIC ...
Page 150: ...150 ARABIC ...
Page 151: ...151 ARABIC ...
Page 152: ...152 ARABIC ...
Page 153: ...153 ARABIC ...
Page 154: ...154 ARABIC ...
Page 155: ...155 ARABIC ...
Page 156: ...156 ARABIC ...
Page 157: ...157 ARABIC ...
Page 158: ...158 ARABIC ...
Page 159: ...159 ARABIC ...
Page 160: ...160 ARABIC ...
Page 161: ...161 ARABIC ...
Page 162: ...162 ARABIC z ...
Page 163: ...163 ...
Page 164: ......