3.5 Data transmission
This oximeter has the function of wireless data transmission. The user could
effectively transmit the data to computer through the wireless communication
module. Refer to the (Oximeter Data Manager) for detailed information.
4. Technical Specifications
A. Technique:
dual-wavelength LED sensor,
LED sensor wavelength:
Red light: 663 nanometers,
Infrared light: 890 nanometers.
Maximal optical output power:
less than 1.5mW maximum average.
B. SpO
2
measurement
Measuring range:
70%~100%
Measuring accuracy:
not greater than 3% for SpO
2
range from 70% to 100%
Note: Accuracy defined as root-mean-square value of deviation
according to ISO 9919.
SpO
2
alarm low limit range:
85%~99% (default 90%)
C. Pulse Rate measurement
Measuring range: 30bpm~240bpm
Measuring accuracy: ±2bpm or ±2% (whichever is greater)
Pulse Rate alarm range:
high limit: 100~240bpm (default 120bpm)
low limit: 30~60bpm (default 50bpm)
D. Perfusion Index (PI) Display
Range: 0.2%~20%
E. Audible &visual alarm function
When measuring, if SpO
2
value or pulse rate value exceeds the preset
alarm limit, the device will alarm automatically and the value which exceeds
limit on the screen will flash. The oximeter will shut down automatically
in 8 seconds with no signal.
F. Display mode:
Color OLED Display
G. Power supply requirement:
2 x LR03 (AAA) alkaline batteries
Supply voltage: 3.0VDC
Operating current: ≤40mA
Summary of Contents for OXY-10
Page 145: ...145 ARABIC ...
Page 146: ...146 ARABIC ...
Page 147: ...147 ARABIC ...
Page 149: ...149 ARABIC ...
Page 150: ...150 ARABIC ...
Page 151: ...151 ARABIC ...
Page 152: ...152 ARABIC ...
Page 153: ...153 ARABIC ...
Page 154: ...154 ARABIC ...
Page 155: ...155 ARABIC ...
Page 156: ...156 ARABIC ...
Page 157: ...157 ARABIC ...
Page 158: ...158 ARABIC ...
Page 159: ...159 ARABIC ...
Page 160: ...160 ARABIC ...
Page 161: ...161 ARABIC ...
Page 162: ...162 ARABIC z ...
Page 163: ...163 ...
Page 164: ......