CORRECTIVE MAINTENANCE
If a malfunction of the equipment is found it is suggested to check first of all the
control setting of the unit.
The unit and its accessories should have checked by professionally qualified personnel
in case that they are exposed:
to external mechanical solicitations, for instance after a serious fall,
to liquid dripping, to strong overheating (for instance direct light of the sun,
fire), or if their functionality seems to have altered or if some part of the unit’s
cabinet it has broken, lacking, or displaced.
or if some connector or cable show signs of deterioration.
PROBLEMS’ SOLUTION
These are some easy solutions to the problems with
IONO BASE PLUS
.
Problem
Solution
The mains switch is switched
on but the unit doesn't work.
Inspect if the main cable is correctly inserted.
Lights are not illuminated.
Check the mains cable connection.
Verify the wall socket functioning.
The unit correctly switches on
but no current has delivered.
Check the lights’colour.
Check the patient’s cable connection.
Light are coloured, the patient
cable has connected correctly,
but the unit doesn't deliver
current.
Check the envelope sponges to be certain they are moist
sufficiently.
Check and replace if need the conductive rubber electrodes.
If the problems are not overcame, please refer to the qualified staff. Do not open the
unit for any reasons, this is dangerous and it will lose every kind of warranty too.