background image

 

 

Operative Manual 

MO082-IT 

22/04/13 

Rev.0 

Pag. 9 di 12 

 

 

 

 

 

6.6

 

Routine maintenance 

 

Period

 

Job 

Once a year

 

Inspect  all  the  lamp  joints  and  make  sure  they  are  not  hard to move.  If Product does not 
maintain the position or is hard in movements, contact the service center. 

Once a year

 

Make sure the retention screws are tightened properly.

 

If these are not properly fastened, 

adequately tighten.

 

Once a year

 

Check the condition of the Product paint. Make sure there are no paint pieces that could 
fall on the operating field.

 

 
 

7.

 

Technical data 

Technical data  

GIMANORD  

Illumination Ec at a distance of 50cm  [Lux] ± 10% 

550 

Data on electrical connection

 

 

Primary alternating voltage [Volt ac] 

220-240 

Frequency [Hz] 

50 

Absorbed power  [VA] 

33 

Light source  

N°1 Lamp. 22W-230V G10q 

Light source duration [h] 
(this  datum  can  vary  according  to  voltage  peaks  and  the 
frequency of use) 

5.000 

General data 

 

Color 

RAL 9003 

Directive  

93/42/EEC (included 2007/47/EC) 

Standard 

EN 60601-1and EN 60601-2-41 

Electrical safety class 

Class I 

Summary of Contents for GIMANORD

Page 1: ...2 1 Installation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 In...

Page 2: ...n The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUIPMENT will be indicated in this manual with the name of Prod...

Page 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Page 4: ...and essential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURAT...

Page 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Page 6: ...ion screws 2 Put the clamp on the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing...

Page 7: ...autions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of s...

Page 8: ...therefore be checked during inspection and maintenance intervals 6 Adjustments 6 1 Yearly inspections by the operator Keep to the yearly inspection schedules and inspect the product according to IEC...

Page 9: ...year Check the condition of the Product paint Make sure there are no paint pieces that could fall on the operating field 7 Technical data Technical data GIMANORD Illumination Ec at a distance of 50cm...

Page 10: ...d indirect contacts B type device Dimensions Overall dimension mm 1200x300x100 Lamp weight Kg 3 Certificate Complying with directive 93 42 EEC included 2007 47 EC All lighting values are subject to a...

Page 11: ...sions IEC 61000 3 3 Compliance Immunity test IEC 60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact...

Page 12: ...nications equipment and the Product The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Product can help...

Reviews: