Operative Manual
MO082-IT
22/04/13
Rev.0
Pag. 8 di 12
5.2
Disinfecting
Switch the Product off by means of the removing of plug and make sure it cannot be switched back on.
Protect the Product against water spray and do not clean it/disinfect it with liquids.
Leave the lamp body to cool down. Only disinfect the lamp body when it is cold.
Disinfectants can contain substances which are harmful for the health: only use disinfectants in accordance with the
rules on hygiene established by the hospital.
The Product operator must comply with the rules established by the national commission for hygiene and
disinfection.
To prevent damaging parts in stainless steel or aluminum, only use disinfectants which are chlorine and halogen
free.
To prevent the plastic parts becoming fragile, use only disinfectants with low alcohol content.
Dose the disinfectants so no liquids penetrate inside the lamp bodies and into the support arm system.
Disinfect the Product with a damp but not wet cloth.
Each Product, over time, is subject to a certain amount of wear. Product safety and operation must therefore
be checked during inspection and maintenance intervals.
6.
Adjustments
6.1
Yearly inspections by the operator
Keep to the yearly inspection schedules and inspect the product according to IEC 62353 standard.
6.2
Repairs
The Product must only be opened and repaired by a technician who has attended a course on the Product organized
by the manufacturer or by a qualified technician in possession of the necessary technical skills.
6.3
Adjustments
The Product is sold already balanced and does not require further adjustment. In the event of the Product becoming
stiff or loose over time, contact the service center.
6.4
Troubleshooting
n
Problem
Solution
1
The Product does not switch on
Check the voltage at input and output of power supply. Substitute
the bulb if there is presence of tension.
2
The Product does not remain in position
Contact the service center.
3
The light flickers
Contact the service center.
6.5
Bulb substitution
After checking that the lamp is switched off, unplugged with cooled bulb and protection screen, it is
possible to unscrew the 4 screws located in the lower part of the aluminum reflector. After that remove
the plexiglass protection screen (2) and substitute the fluorescent circline 22W-230V (3).
Reassemble all the parts as per original state.