2.

 

Product Installation 

Before installing the Product, make sure that all the packages are present and in good condition, without any 
transport-related damage, and that the contents correspond to what is specified above. 
Claims will be considered only if the seller or the forwarder are informed immediately. Each claim must be 
in writing. Goods are always carried at the buyer’s risk.  
Keep the original package in case the lamp must be sent back. 

 

2.1 Installation mobile version 

1.

 

Insert the lamp into the upper hole of the pipe. 

2.

 

Screw down the knob, taking care of the thread 

3.

 

Connect  the  power  cable  plug  to  the  network  line  (please  check  that  the  network  line  is 
equipped with the ground cable). 

 

2.2 Installation wall version (S/12 MED fixing) 

1.

 

Fix the clamp S/12 MED on the wall using 3 expansion screws. 

2.

 

Insert the lamp into the upper hole of the wall clamp S/12MED. 

3.

 

Follow points 2 and 3 of paragraph 

2.1

 

2.3 Installation wall version (rail bar fixing) 

1.

 

Fix the rail bar supports using 2 expansion screws.  

2.

 

Put the clamp on the bar and tighten the handle. 

3.

 

Insert the lamp into the upper hole of the rail clamp 

4.

 

Follow points 2 and 3 of paragraph 

2.1

 

2.4  Installation table version (S11 fixing) 

1.

 

Fix the clamp S/11 to the table tighten the screwed pivot. 

2.

 

Insert the lamp into the upper hole of the table clamp 

3.

 

Screw, using a screwdriver, the screw on the back of the clamp,  

4.

 

Follow point 3 of paragraph 

2.1

 

2.5

 

First starting 

At this point is possible to power the Product to check its correct functioning.  
Follow the steps below: 
1-

 

 Press the green switch of the base; 

 

2.6

 

Installation verification and Product testing operations before use 

The following prescriptions are to be considered compulsory during verification of the installation, since they prove 
the correct execution of all the points set out. For such reason it is necessary to tick each point when it is treated.

 

1.

 

Check the suitability of the wall to the installation of the Product.   

 

2.

 

Check the right installation of the pivot into the clamp/stem 

 

3.

 

Make sure the movement mechanism works perfectly. Check the mechanical operation of the Product 
by directing and rotating

 

 

4.

 

After turning on the Product, it must emit light from the reflector 

 

 
  

    

Seal and signature of the installer:

 

 

 

 
 
 

 

Summary of Contents for GIMANORD

Page 1: ...2 1 Installation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 In...

Page 2: ...n The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUIPMENT will be indicated in this manual with the name of Prod...

Page 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Page 4: ...and essential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURAT...

Page 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Page 6: ...ion screws 2 Put the clamp on the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing...

Page 7: ...autions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of s...

Page 8: ...therefore be checked during inspection and maintenance intervals 6 Adjustments 6 1 Yearly inspections by the operator Keep to the yearly inspection schedules and inspect the product according to IEC...

Page 9: ...year Check the condition of the Product paint Make sure there are no paint pieces that could fall on the operating field 7 Technical data Technical data GIMANORD Illumination Ec at a distance of 50cm...

Page 10: ...d indirect contacts B type device Dimensions Overall dimension mm 1200x300x100 Lamp weight Kg 3 Certificate Complying with directive 93 42 EEC included 2007 47 EC All lighting values are subject to a...

Page 11: ...sions IEC 61000 3 3 Compliance Immunity test IEC 60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact...

Page 12: ...nications equipment and the Product The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Product can help...

Reviews: