background image

 

 

Operative Manual 

MO082-IT 

22/04/13 

Rev.0 

Pag. 12 di 12 

 

 

 

      

Immunity test

 

IEC 60601 test level 

Compliance 

level 

Electromagnetic environment-guidance 

Conducted RF  
IEC 61000-4-6 
 
Radiated RF 
IEC 61000-4-3

 

3 Vrms 
150 kHz to 80 MHz 
 
3 V/m 
80 MHz to 2,5GHz

 

3 Vrms 
 
 
3 V/m

 

Portable and mobile RF communications 
equipment should be used no closer to any part of 
the Product, included cables, than the 
recommended separation distance calculated from 
the equation applicable to the frequency of the 
transmitter. 

Recommended separation distance 

 

d = 1,2√P   

150 KHz to 80 MHz 

d = 1,2√P   

80 MHz to 800 MHz 

d = 2,3√P   

80 MHz to 2,5 GHz 

 

where P is the maximum output power rating of the 
transmitter in watts (W) according to the 
transmitter manufacture and 

d

 is the recommended 

separation distance in meters (m). 
Field strengths from fixed transmitters, as 
determined by an electromagnetic site survey, 
should be less than the compliance leave in each 
frequency range.

 

Interference may occur in the vicinity of equipment 
marked with the following symbol. 

 

     

Recommended separation distance between 

portable an mobile RF communications equipment and the Product 

The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.

 

The 

customer or the user of the Product can help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and the Product as recommended below, 
according to the maximum output power of the communications equipment.

 

Rated maximum output power 

of transmitter 

W

 

Separation distance according to frequency of transmitter 

m

 

 

150 kHz to 80 MHz 

d = 1,2√P   

 

80 MHz to 800 MHz 

d = 1,2√P   

 

800 MHz to 2.5 GHz 

d = 2,3√P   

 

0.01 

0.12 

0.12 

0.24 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in meters (m) 

can be estimated using the equation applicable to the frequency of the transmitter, where 

is the maximum output power 

rating of the transmitter in watts (W) according to the transmitter manufacturer. 

Note 1:

 at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 

Note 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects an people. 

 

Summary of Contents for GIMANORD

Page 1: ...2 1 Installation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 In...

Page 2: ...n The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUIPMENT will be indicated in this manual with the name of Prod...

Page 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Page 4: ...and essential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURAT...

Page 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Page 6: ...ion screws 2 Put the clamp on the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing...

Page 7: ...autions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of s...

Page 8: ...therefore be checked during inspection and maintenance intervals 6 Adjustments 6 1 Yearly inspections by the operator Keep to the yearly inspection schedules and inspect the product according to IEC...

Page 9: ...year Check the condition of the Product paint Make sure there are no paint pieces that could fall on the operating field 7 Technical data Technical data GIMANORD Illumination Ec at a distance of 50cm...

Page 10: ...d indirect contacts B type device Dimensions Overall dimension mm 1200x300x100 Lamp weight Kg 3 Certificate Complying with directive 93 42 EEC included 2007 47 EC All lighting values are subject to a...

Page 11: ...sions IEC 61000 3 3 Compliance Immunity test IEC 60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact...

Page 12: ...nications equipment and the Product The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Product can help...

Reviews: