1.1

 

Qualification of operators 

 

This  paragraph  describes  the  requirements  and  qualifications  which  the  operators  involved  in  the  various  stages  of 
Product life and use must possess. 

 

Installation 

Qualified Installer and/or Technician 

Use 

Professional medical personnel

 

Routine maintenance (about every year) 

Qualified technician in possession of professional technical requirements

 

Special maintenance (in case of necessity)  Qualified technician in possession of professional technical requirements

 

Assistance 

GIMA or authorized dealer 

 

Cleaning 

Properly trained medical and paramedical personnel

 

Demolition  

In compliance with the national directives applicable to waste disposal. 

 

1.2 Packaging, transport, Storage and characteristic of the installation site 

Boxes containing the whole structure, with installation manual and user manual.  
Transport is carried out by any road haulage contractor as long as they respect the following characteristics: 
Temperature (°C): -15 / +60; Humidity: 10 / 75 %; Atmospheric pressure (h/Pa): 500 / 1060. 
The devices packaged must be stored (warehoused) in a dry place and at the following temperature: 
Temperature (°C): +10 / +40;

  

Humidity: 10 / 75 %;

  

Atmospheric pressure (h/Pa): 500 / 1060. 

The site appointed for the installation of the equipment must have the following characteristics:

 

Temperature (°C): +10 / +40;

  

Humidity: 30 / 75 %;

  

Atmospheric pressure (h/Pa): 700 / 1060.

 

 

1.3 Graphic symbols used on the Product and packaging 

Description symbols present on labels, on product, and manual: 

 

 

Graphic  symbol  proving  the  CE  marking 
of the product

 

 

RECYCLING!  Product  must  be  recycled 
separately

 

 

Symbol  indicating  the  manufacture  date 
(month and year) 

 

B-Type  device.  Indicates  the  level  of 
protection  against  direct  and  indirect 
contact

 

 

Consult accompanying document

 

 

Attention 

 

Address of the Manufacturer  

 

Maximum number of packages stackable

 

 

Maximum number of packages stackable

 

 

Breakable package

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for GIMANORD

Page 1: ...2 1 Installation mobile version 6 2 2 Installation wall version S 12MED fixing 6 2 3 Installation wall version rail bar fixing 6 2 4 Installation table version S11 fixing 6 2 5 First starting 6 2 6 In...

Page 2: ...n The ME Medical Electrical EQUIPMENT to which this manual refers is a LAMP for diagnosis or observation For easier description such ME EQUIPMENT will be indicated in this manual with the name of Prod...

Page 3: ...tion manual and user manual Transport is carried out by any road haulage contractor as long as they respect the following characteristics Temperature C 15 60 Humidity 10 75 Atmospheric pressure h Pa 5...

Page 4: ...and essential performance Collateral standard Electromagnetic compatibility Requierements and tests Classification with reference to article 9 and Annex IX of Directives 93 42 EEC and 2007 47 EC DURAT...

Page 5: ...ot above 40 of the net value of the product as indicated on the buyer s invoice GIMA s liability is expressly ruled out for indirect damages or consequential damages including cases of the lamp not be...

Page 6: ...ion screws 2 Put the clamp on the bar and tighten the handle 3 Insert the lamp into the upper hole of the rail clamp 4 Follow points 2 and 3 of paragraph 2 1 2 4 Installation table version S11 fixing...

Page 7: ...autions could cause glare and potential damage to the retina Never place and or hang anything on the Product Unless this precaution is taken positioning will not be reliable and the danger exists of s...

Page 8: ...therefore be checked during inspection and maintenance intervals 6 Adjustments 6 1 Yearly inspections by the operator Keep to the yearly inspection schedules and inspect the product according to IEC...

Page 9: ...year Check the condition of the Product paint Make sure there are no paint pieces that could fall on the operating field 7 Technical data Technical data GIMANORD Illumination Ec at a distance of 50cm...

Page 10: ...d indirect contacts B type device Dimensions Overall dimension mm 1200x300x100 Lamp weight Kg 3 Certificate Complying with directive 93 42 EEC included 2007 47 EC All lighting values are subject to a...

Page 11: ...sions IEC 61000 3 3 Compliance Immunity test IEC 60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact...

Page 12: ...nications equipment and the Product The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Product can help...