GIMA SpA

 

    DIATERMO 

MB122-MB132-MB160 

Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones 

MA036IGBEa  

 

27

 

It is possible to use either single plate neutral reference electrodes or electrodes with split conductive 
zone so to watch the plate to patient contact during the surgical intervention. 
Control of the units is via front panel keys and display; mains inlet is located on the rear panel. 
The units have automatic control systems that, monitoring the internal parameters, signal the possible 
damages/errors that are found.  
The operational parameters that are used are constantly stored so that, every time the unit is switched 
on or the operative method is changed, the last utilized parameters are recalled.  
The level of the emission sound can vary; every operator can choose his own level according to the 
noise of his working ambient. 
Power output can be actived either through holder-handles with pushbuttons or through single or 
double foot switch command. Moreover, applying a special optional adapter it is possible the unit 
connection to bipolar forceps.

 

1.4 MONOPOLAR CUT  

Monopolar cut is the sectioning of the biological tissue achieved by the high-density passage of HF 
current, which is concentrated at point of the active electrode. The HF current, when it is applied to the 
tissue, through the point of the active electrode, it creates intense molecular heat in the cells so high 
that explosion of it is caused.  
The cut effect is achieved by moving the electrode through the tissue and destroying the cells one after 
the other. The movement of the electrode prevents the propagation of the side heat in the tissue, thus 
limiting to a single line the cells’ destruction.  
The best HF current for cutting is pure sine wave without any modulation that cuts very smoothly and 
provides the least thermal effect with poor haemostasis while cutting. Because its effects can be 
precisely controlled, it can be used safely without damage to the bone, but since good coagulation 
while cutting is one of principal benefits of using electrosurgery a current with a certain amount of 
modulation is desirable. 
The following rules help the operator to obtain good cutting, however every user must follow first of all 
his professional judgement as he does every time in his practice. 

 

Keep the tissues moist but not wet 

 

Survey the stroke before activate the electrode 

 

Keep the electrode perpendicular to the tissue 

 

Activate the electrode before making contact with the tissue 

 

Maintain clean the electrode’s tip (the optional sponges F7520 to clean the electrodes are adviced). 

 

Wait at least five seconds before to repeat a stroke. 

 

When the output power is properly set there should be: 

 

no resistance to the electrode movement through the tissue 

 

no change in the cut surfaces colour 

 

no fibres of tissue remained onto the electrode

 

1.5 MONOPOLAR COAGULATION 

Monopolar coagulation is the haemostasis of small blood vessel of the bodily tissue through passing of 
high frequency current in correspondence of active electrode. When the current density is reduced and 
a broad-surfaced electrode is used, to dissipate the energy over a larger area, the effect is to dry out the 
surface cells, without deep penetration, resulting in coagulation. These coagulate surface cells then 
serve as a layer of insulation, preventing heat derived by successive applications of current from 
penetrating too deeply. 
The current normally used for coagulation is modulated and depending from the modulation percentage 
is the smoothness of cutting, goodness of haemostasis and likelihood of tissue destruction. Deeper 
current modulation brings to somewhat roughly cutting and the chance of some slight depth of tissue 
destruction but more efficient coagulation. 
The following rules help the operator to obtain good coagulation: however every user must follow first 
of all his professional judgement as he does every time in his practice. 

 

Select a ball or heavy wire electrode 

Summary of Contents for DIATERMO MB122

Page 1: ...MANUALE D ISTRUZIONI INSTRUCTION S MANUAL MANUAL DE INSTRUCCIONES DIATERMO MB122 MB132 MB160 M30540 MA036IGBEa...

Page 2: ...rsonnel at all times All the safety instructions and advice notes are to be observed Be sure that these operating instructions is furnished together the equipment when this is transferred to other ope...

Page 3: ...D COAG 18 4 4 4 Corrente per Coagulazione Profonda SOFT COAG 18 4 4 5 Corrente di Coagulazione Bipolare BIPOLAR 18 4 5 SEGNALAZIONE DI TEMPO DI EROGAZIONE ECCESSIVO OVT 19 4 6 SEGNALAZIONE DI ECCESSIV...

Page 4: ...on Current SOFT COAG 35 4 4 5 Bipolar Coagulating Current BIPOLAR 35 4 5 SIGNALING OF EXCESSIVE TIME OF DELIVERY OVT 36 4 6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE OC 36 4...

Page 5: ...de la Coagulacion BIPOLAR 53 4 5 EL SE ALAR DE LA POCA EXCESIVA DE LA SALIDA OVT 54 4 6 EL SE ALAR DE LA IMPEDANCIA EXCESIVA EN EL CIRCUITO NEUTRAL DEL ELECTRODO OC 54 4 7 AJUSTE DEL NIVEL AC STICO DE...

Page 6: ...room It has foreseen its use in the monopolar cut cut coagulated or coagulation mode or in bipolar coagulation mode The equipment is conceived for being used in the following sectors DIATERMO descrizi...

Page 7: ...00500 L10 Bent ball electrode 3mm L10 5pcs 5 cm 30529 00500 L10 L Bent ball electrode 3mm L10 L 5pcs 10 cm 30505 00500 L5 Bent hook electrode L5 5pcs 5 cm 30525 00500 L5 L Bent hook electrode L5 L 5pc...

Page 8: ...effects can be precisely controlled it can be used safely without damage to the bone but since good coagulation while cutting is one of principal benefits of using electrosurgery a current with a cert...

Page 9: ...ectioning of the biological tissue due to the passage of HF current concentrated on the tips of the bipolar forceps When the HF current is applied to the tissue between the two tips of the forceps an...

Page 10: ...le or metallic support The use of antistatic sheets is advised Avoid the skin to skin contact for example between arm and body of the patient Insert an interface material like dry surgical gauze Moreo...

Page 11: ...e lack of efficient connection to earth of the installation Operation without a protective earth connection is forbidden Before connect the equipment ascertain that the required voltage showed on the...

Page 12: ...aporate before putting in operation the unit Don t exceed the temperature environment or the allowed moisture Environmental conditions Temperature 10 40 C Relative moisture 30 75 Pressure 70 106k Pa W...

Page 13: ...done pressing the correspondent pushbutton on the unit press the blue foot switch to pre set and deliver the coagulating current the choice between FORCED COAG SOFT COAG or BIPOLAR must be done pressi...

Page 14: ...work 4 1 1 MANUFACTURER S IDENTIFICATION DATA DIATERMO MB122 MB160 and MB132 HF electrosurgical unit are designed manufactured and tested by the LED SpA in its own laboratories in Aprilia LT Italy 4 1...

Page 15: ...ion keyboard 5 Alarm warning for current delivery over time 6 OC open circuit alarm 7 Handle connection 8 Return plate connection MB132 1 Mains switch 2 Pulse time selector 3 Pulse repetition selector...

Page 16: ...t when a deep coagulation together the cut is desired This current is made by sine current e suited for the cut associated to low voltage current suited for coagulation soft coag With this blending a...

Page 17: ...indications 1 Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 2 When the unit has finished to check internal parameters on the display CUT appears the messa...

Page 18: ...push the CONT push button to do it 4 10 2 REPETITION OF TIMED DELIVERY The timed high frequency current emission can be automatically repeated from 2 to 9 times each time one foot switch or one hand s...

Page 19: ...ided with line fuses WARNING before switch on the unit operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net 4 11 2 PEDAL CONNECTOR On the le...

Page 20: ...0 1050 1050 15 Maximum output voltage BLEND Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage FORCED Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage SOFT Vpp on5 2k9 540 540 540 15 Maximum out...

Page 21: ...of the supplier of each item When original reusable accessories supplied by LED S p A are applied the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 C is recommen...

Page 22: ...s properly grounded Assure that all the accessories that should be used are available and sterilised Check by disconnecting the reference electrode cable the functioning of the OC light and sound alar...

Page 23: ...edance load 100 2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nomina...

Page 24: ...2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nominal value Diagrams...

Page 25: ...end of the life the present product mustn t be eliminated as urban refusal but it must be eliminated in a separated collection If the product is eliminated in unsuitable way it is possible that some p...

Reviews: