4.10 MB132 UNITS’ EXTRA FUNCTIONS
This control apply only to the electrosurgical unit model
MB132
that has the following added
characteristics:
Output power delivery time
from 1/1000 up 999/1000 sec
Repetition of delivery actuation
from 2 to 9 times
Session time measaurement
up to 999 min
Output power stroke counter number up to 999
4.10.1
E
MISSION
T
IME
P
RESETTING
If the model
MB132
is used, the actuation time for the delivery of the high
frequency current can be preset with high accuracy. When timed emission is
requested the control of the active time is effected on the frontal panel by
the digitation of the value on the keyboard.
The emission of the high frequency current will occur for the preset time
each time one foot-switch or one hand-switch will be pushed. If, after have
done the timed work, it is necessary to return to the normal work it is
sufficient to push the CONT push button to do it.
4.10.2
R
EPETITION OF
T
IMED
D
ELIVERY
The timed high frequency current emission can be automatically
repeated from 2 to 9 times each time one foot-switch or one hand
switch is pushed. The number of the repetitions, selected through
the pushbutton UP and DOWN, it is pointed out on the
corresponding display. To annul the command of repetition is necessary to press the pushbutton OFF,
in this condition the display has not turned on.
4.10.3
T
REATMENT
T
IME
C
OUNTER
With the
MB132
model it is possible to obtain automatically the elapsed treatment time by presetting
the function by the TIME push-button and starting the counter by the START push-button. To arrest
the counting it is necessary to push the STOP push-button and the RESET push-button is used for the
counter zeroing.
4.10.4
C
OUNTER OF THE
N
UMBER OF
T
IMED
E
MISSION
With the
MB132
model it is possible automatically collect the number of the high frequency current
emission by presetting the function by the PULSE push-button and starting the counter by the START
push-button. To arrest the counting it is necessary to push the STOP push-button and RESET push-
button is used for the counter zeroing.