6. MAINTENANCE 

6.1 GENERAL 

No user adjustable parts are within the equipment, either for calibration or service purposes. The 
equipment housing must not be opened: the warranty is invalidated by any unauthorized entry into the 
unit. In the event any repair or adjustment work being necessary, the whole equipment should be 
returned to the LED S.p.A. Service Centre 04010 APRILIA (LT) - ITALY, or to a other Authorised 
Centre, together with a description of the fault. 
Maintenance work by the user is mainly the cleaning of the exterior of the cabinet, cleaning and 
sterilisation of the accessory items and checking of the equipment before each use. Carrying out function 
and safety check for verification of the parameters is demanded to specialised technical people. 

6.2 CLEANING OF THE CABINET 

Switch the equipment off completely and disconnect the mains supply before any cleaning is 
undertaken. Clean the outside of the cabinet with a damp cloth. No chemical should be used; a mild 
non abrasive cleanser may be used when necessary.  

6.3 CLEANING AND STERILISATION OF THE ACCESSORY ITEMS 

The best thing to do is to use only one time use accessories and discard them after use. Since some of 
the accessory items are to be used more than once it is mandatory to clean carefully and sterilise those 
accessories before the new use. The best way to clean and sterilise the reusable items is to follow the 
direction of the supplier of each item. When original reusable accessories supplied by LED S.p.A. are 
applied, the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 °C is 
recommended. 

6.4 GUIDE TO THE SOLUTION OF THE PROBLEMS 

In case of problems before all it is advised to check for the correct installation of the unit and for the 
correct connection of the accessories.   

Problems Probable 

Cause 

Solution 

The equipment 
doesn't switch on. 

Interruption or absence of the  main 
feeding 

Verify the connection of the main cable. 
Verify the fuses and replace them, where 
necessary, with new ones of the propre type. 

Alarm OC always 
active. 

Interruption or lack of  contact on 
the neutral electrode circuit 

Check the connection of the cable to the 
neutral electrode. Replace the cable of 
connection of the neutral electrode.  

The unit doesn't 
respond to the 
command of 
activation. 

Breakdown of the handpiece or of 
the pedal - Wrong connection of the 
handpiece or of the pedal - Alarm 
OVT activated 

Replace the handpiece or the pedal.  
Verify the connection of the handpiece or of 
the pedal. 
Wait for the OVT warning signal getting out. 

Error Code 001 

Current delivery control activated 
during switching on  

Disconnect the handpiece or the pedal and 
switch on the unit again. 

Error Code 002 

Error in the management board   

Call for Service   

Error Code 003 

Error in the management board   

Call for Service   

Error Code 004 

Error in the data conversion circuit    Call for Service   

Error Code 005 

Error of the  reference voltage value    Verify the main voltage. Call for Service   

Error Code 009 

Error in the output  power activation 
circuit    

Call for Service   

Error Code 010 

Error in the output  power activation 
circuit  

Call for Service   

Summary of Contents for DIATERMO MB122

Page 1: ...MANUALE D ISTRUZIONI INSTRUCTION S MANUAL MANUAL DE INSTRUCCIONES DIATERMO MB122 MB132 MB160 M30540 MA036IGBEa...

Page 2: ...rsonnel at all times All the safety instructions and advice notes are to be observed Be sure that these operating instructions is furnished together the equipment when this is transferred to other ope...

Page 3: ...D COAG 18 4 4 4 Corrente per Coagulazione Profonda SOFT COAG 18 4 4 5 Corrente di Coagulazione Bipolare BIPOLAR 18 4 5 SEGNALAZIONE DI TEMPO DI EROGAZIONE ECCESSIVO OVT 19 4 6 SEGNALAZIONE DI ECCESSIV...

Page 4: ...on Current SOFT COAG 35 4 4 5 Bipolar Coagulating Current BIPOLAR 35 4 5 SIGNALING OF EXCESSIVE TIME OF DELIVERY OVT 36 4 6 SIGNALING OF EXCESSIVE IMPEDANCE IN THE CIRCUIT OF NEUTRAL ELECTRODE OC 36 4...

Page 5: ...de la Coagulacion BIPOLAR 53 4 5 EL SE ALAR DE LA POCA EXCESIVA DE LA SALIDA OVT 54 4 6 EL SE ALAR DE LA IMPEDANCIA EXCESIVA EN EL CIRCUITO NEUTRAL DEL ELECTRODO OC 54 4 7 AJUSTE DEL NIVEL AC STICO DE...

Page 6: ...room It has foreseen its use in the monopolar cut cut coagulated or coagulation mode or in bipolar coagulation mode The equipment is conceived for being used in the following sectors DIATERMO descrizi...

Page 7: ...00500 L10 Bent ball electrode 3mm L10 5pcs 5 cm 30529 00500 L10 L Bent ball electrode 3mm L10 L 5pcs 10 cm 30505 00500 L5 Bent hook electrode L5 5pcs 5 cm 30525 00500 L5 L Bent hook electrode L5 L 5pc...

Page 8: ...effects can be precisely controlled it can be used safely without damage to the bone but since good coagulation while cutting is one of principal benefits of using electrosurgery a current with a cert...

Page 9: ...ectioning of the biological tissue due to the passage of HF current concentrated on the tips of the bipolar forceps When the HF current is applied to the tissue between the two tips of the forceps an...

Page 10: ...le or metallic support The use of antistatic sheets is advised Avoid the skin to skin contact for example between arm and body of the patient Insert an interface material like dry surgical gauze Moreo...

Page 11: ...e lack of efficient connection to earth of the installation Operation without a protective earth connection is forbidden Before connect the equipment ascertain that the required voltage showed on the...

Page 12: ...aporate before putting in operation the unit Don t exceed the temperature environment or the allowed moisture Environmental conditions Temperature 10 40 C Relative moisture 30 75 Pressure 70 106k Pa W...

Page 13: ...done pressing the correspondent pushbutton on the unit press the blue foot switch to pre set and deliver the coagulating current the choice between FORCED COAG SOFT COAG or BIPOLAR must be done pressi...

Page 14: ...work 4 1 1 MANUFACTURER S IDENTIFICATION DATA DIATERMO MB122 MB160 and MB132 HF electrosurgical unit are designed manufactured and tested by the LED SpA in its own laboratories in Aprilia LT Italy 4 1...

Page 15: ...ion keyboard 5 Alarm warning for current delivery over time 6 OC open circuit alarm 7 Handle connection 8 Return plate connection MB132 1 Mains switch 2 Pulse time selector 3 Pulse repetition selector...

Page 16: ...t when a deep coagulation together the cut is desired This current is made by sine current e suited for the cut associated to low voltage current suited for coagulation soft coag With this blending a...

Page 17: ...indications 1 Switch on the unit through the mains switch while the CUT pushbutton is maintained pressed 2 When the unit has finished to check internal parameters on the display CUT appears the messa...

Page 18: ...push the CONT push button to do it 4 10 2 REPETITION OF TIMED DELIVERY The timed high frequency current emission can be automatically repeated from 2 to 9 times each time one foot switch or one hand s...

Page 19: ...ided with line fuses WARNING before switch on the unit operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net 4 11 2 PEDAL CONNECTOR On the le...

Page 20: ...0 1050 1050 15 Maximum output voltage BLEND Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage FORCED Vpp on5 2k9 1050 1050 1050 15 Maximum output voltage SOFT Vpp on5 2k9 540 540 540 15 Maximum out...

Page 21: ...of the supplier of each item When original reusable accessories supplied by LED S p A are applied the cleaning by using soft cleanser and sterilization through steam sterilisation at 121 C is recommen...

Page 22: ...s properly grounded Assure that all the accessories that should be used are available and sterilised Check by disconnecting the reference electrode cable the functioning of the OC light and sound alar...

Page 23: ...edance load 100 2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nomina...

Page 24: ...2000 SOFT Diagrams of half and maximum output power versus impedance load 10 1000 BIPOLAR Diagrams of output power CUT versus nominal value Diagrams of output power BLEND versus nominal value Diagrams...

Page 25: ...end of the life the present product mustn t be eliminated as urban refusal but it must be eliminated in a separated collection If the product is eliminated in unsuitable way it is possible that some p...

Reviews: