ACCESSORIES SUPPLIED 

 

DESCRIPTION 

COMPLETE ASPIRATION JAR 1000cc 

CONICAL FITTING 

TUBES SET 8 mm x 14 mm 

ANTIBACTERIAL FILTER 

AC/DC ADAPTER 

EUROPEAN POWER SUPPLY CORD FOR 

AC/DC ADAPTER 

CIGARETTE LIGTHER CABLE 

 
 

The filter is produced with (PTFE) hydrophobic material to prevent fluids entering the pneumatic circuit. It should be changed 
immediately if it becomes wet or if there is any sign of contamination or discolouration. If should also be changed if the unit is used with 
a patient whose risk of contamination is unknown. 

Don’t use the suction unit without the protection filter

. If the suction unit is used in 

an emergency or in a patient where the risk of contamination is not know the filter must be changed after each use.  
Available under request with different versions with complete jar 2000cc.  

 
WARNING: 

The medical device is provided without a specific suction probe.  If this device must be used with a specific suction probe, 

the end user is responsible for making sure it complies with the EN 10079-1 regulation. 
 

WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 
standards on material biocompatibility. 

 

CLEANING OF ACCESSORIES 

 

To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions. 
Washing and / or cleaning the autoclavable jar as to be carried out as follows: 

 

Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances; 

 

Disconnect the jar from the device 

 

Disconnect all tubes from the jar and the protection filter 

 

Empty and dispose of the content and of the suction catheter according to the laws in force in your country; 

 

Separate all parts of the cover (overflow valve, o-ring); 

After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly.                                                         
Then soak in warm water (temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to remove 
incrustations. Rinse in running warm water and dry all parts with a soft cloth (non-abrasive). 
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar 
relative pressure – 15min) making sure that the jar is positioned upsidedown. 
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1).                             
Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore 
recommended. 
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged. 
Assemble the jar as follows: 

 

Place the overflow valve into its seat in the cover (under VACUUM connector) 

 

Insert floating valve keeping the o-ring towards the opening of the cage 

 

Place the o-ring into its seat around the cover 

 

After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid 
exit 

 

The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C (1 bar relative pressure – 15min. The conical 
connector can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15min.   
The device is ready for a new employment now.  
 

 

DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER

 

 
 
 
 
 
 
 

Summary of Contents for 28190

Page 1: ...MANUALE D USO INSTRUCTION MANUAL MODE D EMPLOI HANDBUCH MANUAL DE ISTRUCCIONES GIMA S p A Via Marconi 1 20060 Gessate MI Italia ITALIA Tel 02 953854 1 Fax 02 95381167 E mail gima gimaitaly com www gi...

Page 2: ...fabbricante al fine di garantire la massima efficienza e sicurezza del dispositivo Utilizzare il dispositivo medico sempre con il filtro antibatterico in dotazione Non immergere mai l apparecchio in a...

Page 3: ...a prevenire potenziali conseguenze negative per l ambiente e per la salute che potrebbero altrimenti essere causate dal loro inadeguato smaltimento Il riciclaggio dei materiali aiuta a conservare le r...

Page 4: ...tura di immagazzinamento 40 70 C Apparecchio di tipo B Corrente Alternata Corrente Continua Batteria Pb Battery 12V 4A Classe di Isolamento II solo se collegato con cavetto accendisigari per auto Hz F...

Page 5: ...arare tutte le parti del coperchio dispositivo troppo pieno guarnizione Smaltite le parti monouso e disassemblato il flacone immergere in acqua fredda corrente e risciacquare bene In seguito immergere...

Page 6: ...e IEC EN 61000 3 2 Classe A Fluttuazioni di tensione flicker IEC EN 61000 3 3 Conforme Guida e dichiarazione del costruttore immunit elettromagnetica L aspiratore chirurgico SUPER VEGA BATTERY utilizz...

Page 7: ...essere minore del livello di conformit di ciascun intervallo di frequenza b Si pu verificare interferenza in prossimit di apparecchi contrassegnati dal seguente simbolo Immunit Irradiate IEC EN 61000...

Page 8: ...RSI CHE LE PARTI INTERNE DELL APPARECCHIO NON VENGANO A CONTATTO CON LIQUIDI E CHE ALIMENTATORE O BATTERIA SIANO DISINSERITI PRIMA DI PROCEDERE ALLA PULIZIA MODALITA DI RIENTRO PER RIPARAZIONE NEL RIS...

Page 9: ...o 4 Mancata aspirazione Guarnizione del coperchio non in sede Svitare il coperchio e riposizionare la guarnizione nella sede del coperchio 5 Galleggiante bloccato Incrostazioni sul galleggiante Svitar...

Page 10: ...ale al dispositivo tramite apposito connettore ed inserire la spina del cavo di alimentazione alla presa di corrente Per iniziare il trattamento premere l interruttore sulla posizione I per accendere...

Page 11: ...vo in funzione dopo aver premuto il pulsante di Accensione il LED resta acceso VERDE FISSO Segnalazione Led Fase Problema Causa Soluzione Led verde lampeggiante Durante la Carica Carica della batteria...

Page 12: ...d on the adapters and extensions 4 Pay special attention to the following Only use original accessories and components The device can be used only with the bacteriological filter Never immerge the app...

Page 13: ...er to point out the obligation to dispose of the electromedical appliances separately the product has the sign of the crossed mobile waste container applied to it TECHNICAL CHARACTERISTICS Model SUPER...

Page 14: ...20060 Gessate MI Italy Warning consult the instruction manual Keep in a cool dry place Conservation temperature 40 70 C Type B equipment Alternate Current Battery Direct Current Insulation Class II on...

Page 15: ...protection filter Empty and dispose of the content and of the suction catheter according to the laws in force in your country Separate all parts of the cover overflow valve o ring After disposing of...

Page 16: ...istribution that supplies power to environments used for domestic scopes or environments feeds to you from batteries Harmonic emissions IEC EN 61000 3 2 Class A Voltage fluctuations flicker emissions...

Page 17: ...ers and TV transmitters can not be theoretically and accurately foreseen To establish an electro magnetic environment generated by fixed RF transmitters an electro magnetic study of the site should be...

Page 18: ...doesn t close If the cap has been washed ensure that the float is not partially detached Insert the float into it s place 6 The float doesn t close The float it s covered by dirty material Unscrewed t...

Page 19: ...n rescrew the lid on the jar correctly To extract the accessories and start with cleaning WARNING Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into t...

Page 20: ...GIMA will substitute the instrument only if a SALE RECEIPT and STAMPED GUARANTEE accompany the same GIMA is not responsable for contaminated accessories they will be substitute at customer s expenses...

Page 21: ...curit indiqu es pour les appareils lectriques et notamment Utiliser seulement les accessoires et les composants originaux Ne jamais plonger l appareil dans l eau Positionner l appareil sur des surfac...

Page 22: ...MAXIMUM sans bocal 36 l min POIDS 4 39 kg CLASSE D ISOLEMENT si utilis avec une alimentation Courant alternatif Courant continu mod UE60 140429SPA3 Classe I CLASSE D ISOLEMENT si utilis avec batterie...

Page 23: ...notice d utilisation Conserver dans un local frais et sec Temp rature de stokage 40 70 C Appareil de type B Fusible Courant alternatif Hz Fr quence du secteur Corrente Continua Batterie Batterie au P...

Page 24: ...our des raisons domestiques or alimentation des environnements vous des batteries Harmoniques IEC EN 61000 3 2 Classe A Fluctuations de tension flicker IEC EN 61000 3 3 Conforme Guide et declaration d...

Page 25: ...re pr vus th oriquement et avec pr cision Pour tablir un environnement lectromagn tique caus par des transmetteurs RF fixes il faut consid rer la r alisation d une investigation lectromagn tique du si...

Page 26: ...cle dispositif trop plein garniture Une fois qu on a limin les parties jetables et on a d sassembl le flacon il faut plonger les parties dans l eau froide courent et les rincer fond Ensuite il faut pl...

Page 27: ...eur L appareil se compose d une batterie au plomb qui ne peut pas tre atteinte de l ext rieur Pour la remplacer consulter le personnel technique autoris par le fabricant D faut Type Cause Solution 1 V...

Page 28: ...ant la manette dans le sens des aiguilles d une montre on obtient une valeur de d pression sup rieure les valeurs en question peuvent tre lues sur l instrument appel vacuom tre Pour suspendre et ou te...

Page 29: ...ES QUI OPERENT POUR OBTENIR LA QUALITE ET LE BIEN ETRE Tout appareil qui sera restitu GIMA sera soumis des contr les d hygi ne avant la r paration Si GIMA jugera l appareil irr parable cause de signes...

Page 30: ...e verstopft werden Das Ger t nicht in R umen benutzen in denen brennbare An sthesiegemische mit Luft mit Sauerstoff oder mit Lachgas vorhanden sind Das Ger t nicht mit feuchten H nden anfassen Auf jed...

Page 31: ...25 bar Max Saugleistung nie Flasche 36 l min Gewicht 4 39 kg Isolierstoffklasse wenn mit AC DC Netzger t Mod UE60 140429SPA3 verwendet Klasse I Isolierstoffklasse wenn mit interner Batterie verwendet...

Page 32: ...IE KORREKTE ENTSORGUNG DES PRODUKTS IN BEREINSTIMMUNG MIT DER EGRICHTLINE 2002 96 EG Am Ende siner Nutzzeit darf das Produkt NICHT zusammen mit dem Siedlungsabfall beseitigt werden Es kann zu den von...

Page 33: ...n geeignet einschlie lich f r h uslichen Gebrauch und f r direkten Anshluss An den Haushaltstrom Oberschwingungen IEC EN 61000 3 2 Klasse A Spannungsschwankungen Flimmern IEC EN 61000 3 3 Konform Anle...

Page 34: ...nen theoretisch und mit Genauigkeit nicht vorausgesehen werden Um ein elektromagnetisches Umfeld festzulegen das durch feste RF Sender verursacht wird m sste eine elektromagnetische Untersuchung des O...

Page 35: ...Sie alle Schl uche von dem Beh lter und dem Schutzfilter Leeren und entsorgen Sie den Inhalt und den Absaugkatheter gem den in Ihrem Land geltenden Bestimmung Entnehmen Sie alle Teile der Abdeckung b...

Page 36: ...ngen auszutauschen muss man sich an das autorisierte technische Personal des Herstellers wenden Die Bleibatterie des Ger ts ist nicht von au er zug nglich F r den Austausch der Batterie wenden Sie sic...

Page 37: ...n Beh lter zu 1 3 mit Wasser f llen um das Reinigen und den Vakuumaufbau w hrend des Betriebs zu beschleunigen Den Deckel dann auf den Beh lter aufschrauben Die Zubeh rteile herausnehmen und reinigen...

Page 38: ...gger te SUPER VEGA BATTERYeine Garantie con 24 Monaten ab dem Kaufdatum F r die G ltigkeit der Garantie mu der Kunde folgende Dokumente bereit stellen Vorlage der Rechnungskopie und oder Kaufbest tigu...

Page 39: ...la toma el ctrica Efectuar dichos controles antes de cada uso 2 Antes de conectar el aparato comprobar siempre que los datos el ctricos indicados en la etiqueta de los datos y el tipo de enchufe util...

Page 40: ...s TRATAMIENTO DE LAS BATER AS AL FINAL DE SU VIDA TIL Directiva 2006 66 CE Este s mbolo en el producto indica que las bater as no pueden ser tratadas como un residuo dom stico normal Al asegurarse de...

Page 41: ...de los transductores y cables vendidos por el fabricante del aparato y del sistema como piezas sustituibles puede resultar en un incremento de las emisiones o en una disminuci n de la inmunidad del di...

Page 42: ...ectromagn tica El aspirador SUPER VEGA BATTERYse puede utilizar en el ambiente electromagn tico que se especifica a continuaci n El cliente y o el usuario del aspirador SUPER VEGA BATTERYdeben asegura...

Page 43: ...venir interferencias electromagn ticas asegurando una distancia m nima entre aparatos de comunicaci n m viles y port tiles de RF transmisores y el aparato SUPER VEGA BATTERYcomo se indica a continuaci...

Page 44: ...jar en agua medio caliente la temperatura no debe exceder del 60 C Lavar en profondo y si se requiere usar un cepillo non abrasivo para quitar cualquier incrustaci n Aclarar en agua caliente corrente...

Page 45: ...iminar por separado en el produco aparece un contenedor de basura m vil tachado CONTROL PERI DICO DE MANTENIMIENTO El aparato SUPER VEGA BATTERYno tiene piezas que necesiten mantenimiento y o lubricac...

Page 46: ...ura de la tapa del vaso en cuyo interior est montado el flotante Cuando se alcanza el 90 del volumen del frasco se activa el flotador de seguridad el flotador cierra el connector de aspiraci n del fra...

Page 47: ...al en la posici n 0 La autonom a de la bater a con carga plena es de aproximadamente 60 minutos con funcionamiento continuado TAB I INDICACIONES LUMINOSAS DURANTE EL FUNCIONAMIENTO Con alimentaci n ex...

Page 48: ...coni 1 20060 Gessate MI Italia ITALIA Tel 02 953854 1 Fax 02 95381167 e mail gima gimaitaly com www gimaitaly com INTERNATIONAL Tel 39 02 953854209 221 225 Fax 39 02 95380056 e mail export gimaitaly c...

Reviews: