Chapter 2 
GENERAL

2.1 Function and application

The nebulizer can atomize medication into a mist of microscopic droplets, which can be easily inhaled into respiratory system 

along with breathing, achieving therapeutic effect for respiratory diseases such as acute inflammation of the upper respiratory 

tract, acute and chronic tracheitis, bronchitis and swelling and pain in throat, etc. 

Contraindications:

The patients with bronchial asthma or acute emphysema are not recommended to use the device, please follow the doctor's 

advice.

2.2 Features

Power supply: DC 5 V or 2 “AA” alkaline batteries

Input power: <3 VA

Nebulization rate: ≥0.25 mL/min

Noise: ≤50 dB

Equivalent volume particle diameter distribution: the occupation of small atomized particles (diameter < 5 μm) is no less than 

90%.

Type of protection against electric shock: Class II

Degree of protection against electric shock: type BF applied part

Degree of protection against ingress of liquid: IP22

Note: please choose the power adapters manufactured by qualified companies(input: AC100-240V, 50Hz / 60Hz, output: DC5V, 

1A).

The voltage of 2 “AA” alkaline batteries is DC3V.

2.3 Operational environment

Temperature: 5°C~40°C

Humidity: 15%~90%

Atmospheric pressure: 700hPa~1060hPa

Attention: This product is not suitable for use in strong electromagnetic interference environments (such as various medium/

high frequency therapeutic instruments, transformers, large electrical cabinets, radio and television transmission towers, other 

radio frequency transmitting equipment, and other electrical appliances or medical equipment which may generate interference).

2.4 Principles

Principle of nebulization

The high-frequency vibration of piezoelectric ceramic plate directly leads to the deformation of microplate, making medication 

in contact with the microplate extruded from it to form a mist spray effect. The mesh nebulizer is applicable for use in hospital, 

clinic and family.

Principle of treatment

Respiratory system is an open system. The atomized medication, after inhalation, can be directly adsorbed on patient's oral 

cavity, throat, trachea, bronchus and pulmonary alveoli, etc., through its mucous membrane absorption to achieve the purpose 

of treatment.

Chapter 3 
PRODUCT COMPOSITION

Component description: 

The nebulizer consists of main unit, medi-

cation  cup,  mask  and  the  power  adapters 

(optional).

Nebulizer:

Mouthpiece 

Electrode 

ON/OFF Button

Medication cup

Main unit

Summary of Contents for 28066

Page 1: ...TTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produ...

Page 2: ...thetic mixture Otherwise it may cause personal injury Do not wrap the power cord around children s neck or other sensitive areas or it may cause asphyxia The mask of this equipment is made of PVC mate...

Page 3: ...batteries is DC3V 2 3 Operational environment Temperature 5 C 40 C Humidity 15 90 Atmospheric pressure 700hPa 1060hPa Attention This product is not suitable for use in strong electromagnetic interfer...

Page 4: ...cup to the main unit be sure to install it properly until clasp sound is heard Otherwise it may cause electrode conduction failure then the device cannot atomize normally 2 Remove medication cup from...

Page 5: ...ery life will be affected Remove the batteries if the device won t be used for long time 4 2 Operations for treatment use Preparations before use 1 Remove medication cup please clean and disin fect it...

Page 6: ...the cup lid and discard residual medication 3 Add 75 ethanol solution in medication cup cover the cup lid then leave for at least 10mins it is advisable to gently shake it for better disinfection 4 Im...

Page 7: ...medication cup Trickle medication into medication cup re member not to exceed its maximum capacity Improper medication Consult a doctor about whether the medica tion is suitable for the device The ne...

Page 8: ...ode Keep away from sunlight 55 C 40 C Store between 40 and 55 C Lot number Fragile handle with care Date of manufacture Serial number Follow instructions for use Medical Device complies with Directive...

Page 9: ...4 2 Air Discharge 2 kV 4 kV 8 kV 15 kV contact 8 kV air Air Discharge 2 kV 4kV 8 kV 15 kV contact 8 kV air Floors should be wood concrete or ce ramic tile If floor is covered with synthetic material...

Page 10: ...ct was purchased If the equipment is not disposed of correctly fines or penalties may be applied in accordance with the national legislation and regulations GIMA WARRANTY CONDITIONS Congratulations fo...

Reviews: