User Manual AWR Nebulizer MN-300/2 

 

 

Summary of Contents for MN-300/2

Page 1: ...User Manual AWR Nebulizer MN 300 2 ...

Page 2: ......

Page 3: ...stributing Company AWR Inhalation GmbH Am Stadtweg 48 B 15834 Rangsdorf Germany Email info awr inhalation de Tel 49 33708 355020 Online http www awr inhalation de Document version MN 300 2 EN V3 NOTE Please carefully read the user manual before first use of the AWR Nebulizer MN 300 2 Please store the user manual in a safe place ...

Page 4: ...D Displays and Graphical Symbols 10 4 Intended Use 12 5 Cleaning 17 6 Power Supply 18 7 Environmental Conditions 20 8 Disposal 21 9 Consumables 22 10 Maintenance Service and Order information 23 11 Technical specifications 24 12 Performance Characteristics and Aerosol Spectrum 25 13 Information on Electromagnetic compatibility 26 14 EC Declaration of Conformity 30 15 Warranty 30 ...

Page 5: ...r Never attach moist or wet nebulizer units to the control unit make sure the nebulizer unit is dry before every use 2 The device is unsuitable for use in proximity to combustible gases 3 The device must not come into contact with heated surfaces 4 The AWR Nebulizer MN 300 2 is not an officially certified medicine product and therefore not approved for nebulizing medications Usage and choice of po...

Page 6: ...If an air bubble forms behind the metal membrane or if no liquid touches the membrane the control unit will switch off automatically 4 Over the course of their uses the nebulizer units can change their technical properties technical wear and tear This can negatively affect the properties of the produced aerosol Furthermore the nebulizer units should be replaced regularly for reasons of hygiene Dep...

Page 7: ...ial state the charging time for the internal rechargeable battery lithium ion battery of the control unit is about 2 hours for 80 charging capacity and about 3 hours for 100 charging capacity 5 This product is subject to the Federal German law concerning distribution take back and environmentally sound disposal of electric and electronic devices Elektro and Elektronikgesetz ElektroG For this reaso...

Page 8: ...leteness before first operation of the AWR Nebulizer MN 300 2 Also inspect the delivered components for visual defects such as contamination stress cracks or damage The AWR Nebulizer MN 300 2 or its components must not be used if they display visual defects or damage Contact the distributing company in such cases The following table lists all components of the AWR Nebulizer MN 300 2 Article Descri...

Page 9: ...cription Image 201 102 05 Power adapter AWR Nebulizer MN 300 2 201 102 02 Optionally available USB charging cable AWR Nebulizer MN 300 2 201 102 06 Optionally available Nebulizer unit AWR Nebulizer MN 300 2 MMD 6 0 µm 201 102 07 Optionally available Mouth piece AWR Nebulizer MN 300 2 ...

Page 10: ...Air humidity limit Indicates the safe air humidity range for the product LED displays of the control unit Symbol LED display Meaning Battery status indicator The LED shows a green light when operating the device with battery capacity over 30 switches between green and red if the battery capacity is above 10 and below 30 shows a red light if the battery capacity is below 10 shows a slow green flash...

Page 11: ...as responsible for the conformity assessment procedure Article number Shows the order number designated by the distributing company in order to identify the article Type B applied part Device of protection class II Observe the instructions of the user manual Reiterates the necessity for users to refer to the user manual Attention Reiterates the necessity for users to look through to the user manua...

Page 12: ... have to be approved by the manufacturers pharmaceutical companies for use with this type of application inhalation via ultrasonic nebulizing Step 1 Filling the nebulizer unit The liquid recommended for inhalative lung care is a 0 9 saline solution Open the sealing cap by turning it 80 degrees counter clockwise and then lifting it up Fill the liquid into the chamber Re attach the sealing cap and c...

Page 13: ... with the male USB connector of the nebulizer unit Step 3 Starting the Control Unit Press the on off button of the control unit for a prolonged period 1 second The control unit starts up This is indicated by a short acoustic signal and the battery status indicator shows a continuous green light Operational readiness of the AWR Nebulizer MN 300 2 is provided when the operating mode indicator has sh...

Page 14: ... the manual mode When you switch on the AWR Nebulizer MN 300 2 and it s ready for operation the AWR Nebulizer MN 300 2 will be in manual mode You can perform the switch from manual to continuous mode within a timeframe of 60 seconds after startup or last manual nebulizing This corresponds to the maximum standby time Please now press the on off button of the control unit briefly 1 second You will h...

Page 15: ...e it from your mouth when you breathe out Repeat this process until the filled liquid has fully been aerosolized The control unit will then switch off automatically NOTE During each production phase for aerosol the operating mode indicator lights up blue for the whole duration of the aerosol production ATTENTION Please make sure to always hold the AWR Nebulizer MN 300 2 during nebulizing in a way ...

Page 16: ...off mechanisms These are the automatic switch off and the maximal runtimes of the AWR Nebulizer MN 300 2 The automatic switch off switches the AWR Nebulizer MN 300 2 off automatically regardless of the chosen nebulizing mode as soon as the filled liquid has fully been nebulized The maximal runtimes switch off the AWR Nebulizer MN 300 2 automatically as soon as the respective maximal durations have...

Page 17: ...here refers to both the AWR Nebulizer MN 300 2 nebulizer unit with an MMD of 4 0 µm and the optionally available AWR Nebulizer MN 300 2 nebulizer unit with an MMD of 6 0 µm Clean the nebulizer unit by filling a small amount 0 5 ml to 2 0 ml max of distilled or sterile water into the chamber and have the device nebulize it in continuous mode WARNING Do not inhale the nebulized distilled or sterile ...

Page 18: ...battery capacity falls below the threshold of 10 The AWR Nebulizer MN 300 2 will switch off automatically The battery has to be recharged if the battery capacity falls below the threshold of 10 ATTENTION The charging process for the internal rechargeable battery lithium ion battery must only be performed with the provided power adapter or the optionally available USB charging cable to avoid damagi...

Page 19: ...n 1 second pause NOTE Depending on the initial state the charging time for the internal rechargeable battery lithium ion battery of the control unit for the AWR Nebulizer MN 300 2 is about 2 hours for 80 charging capacity and about 3 hours for 100 charging capacity ATTENTION The internal rechargeable battery lithium ion battery of the AWR Nebulizer MN 300 2 will only reach full potential after 2 3...

Page 20: ...ons Reliable operation of the device is guaranteed under the following environmental conditions Ambient air temperature 10 C to 40 C or 50 F to 104 F Relative air humidity 25 to 75 Atmospheric pressure 450 hPa to 1100 hPa Shipping and storage conditions Shipping and storage of the device are guaranteed under the following environmental conditions Ambient air temperature 10 C to 50 C or 14 F to 122...

Page 21: ... ElektroG For this reason the electronic components of the AWR Nebulizer MN 300 2 control unit and power adaptor must not be disposed as domestic waste Disposal of these electronic components has to meet local disposal policies Subject to the German waste catalogue ordinance Abfallverzeichnis Verordnung AVV waste code 20 01 39 separately collected fractions synthetic materials the nebulizer units ...

Page 22: ...son should use their own personal nebulizer unit as well as a personal mouth piece ATTENTION Over the course of their uses the nebulizer units can change their technical properties technical wear and tear This can negatively affect the properties of the produced aerosol Furthermore the nebulizer units should be replaced regularly for reasons of hygiene Depending on use frequency the nebulizer unit...

Page 23: ...rmed by the distributing company Failure to heed this information may lead to loss of warranty Please contact the distributing company if you need technical service or advice Replacement parts and consumables especially the nebulizer units with an MMN of 4 0 µm and 6 0 µm which have to be replaced regularly can be obtained via the online shop of the distributing company or specialist retailers AWR...

Page 24: ...ting frequency 110 kHz Nebulizer capacity 0 6 ml min MMAD 4 0 µm Electrical protection class Class II type B Type of battery used Lithium ion battery 3 7 V 1200 mAh Power adapter AWR Nebulizer MN 300 2 Size 74 x 44 x 73 mm Cable length 1 8 m Weight 165 g Nebulizer unit MMD 4 0 µm AWR Nebulizer MN 300 2 Size 57 x 56 x 30 mm Maximum filling capacity 8 ml Weight 20 g Nebulizer unit MMD 6 0 µm AWR Neb...

Page 25: ... EN 13544 1 appendix CC CC 2 is 1 86 mg sodium chloride per minute This is equivalent to an aerosol delivery of 0 4 ml sodium chloride per minute The average particle size as per DIN EN 13544 1 appendix CC is 3 68 μm MMD with a GSD of 1 64 Performance characteristics and aerosol spectrum for the AWR Nebulizer MN 300 2 when using the nebulizer unit with an MMD of 6 0 µm Aerosol delivery as per DIN ...

Page 26: ...y for its internal FUNCTION Thus its HF emission is very low and electromagnetic interferences with electronic devices in the proximity are unlikely HF emission as per CISPR 11 Class B The nebulizer is suitable for use in any type of facilities including living quarters and those that are directly connected to a PUBLIC SUPPLY NETWORK which also supplies buildings that are used for residential purp...

Page 27: ...ernal conductor to external conductor 2 kV external conductor to earth 1 kV voltage external conductor to external conductor 2 kV external conductor to earth The quality of the supply voltage should be to the standard of a typical business or hospital environment Voltage dips short interruptions and voltage variations as per IEC 61000 4 11 5 UT 95 voltage dip of UT for period 40 UT 60 voltage dip ...

Page 28: ...hould be less than the compliance levelb in each frequency range Interference may occur in the vicinity of equipment marked with NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations The emanation of electromagnetic variables is affected by absorption and reflection by buildings objects and persons a Field strengths from fixed RF tr...

Page 29: ... Output power of transmitter in W Safety distance depending om the transmitted frequency 150 kHz to 80 MHz D 1 2 P 80 MHz to 800 MHz D 1 2 P 800 MHz to 2 5 GHz D 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters whose maximum power output is not covered by the table above the recommended safety distance D in meters m can be calculated using the...

Page 30: ...ed for use within certain voltage limits 15 Warranty We grant you a 24 months warranty starting on the date of the purchase for your AWR Nebulizer MN 300 2 This warranty applies to the control unit of the AWR Nebulizer MN 300 2 and power adapter of the AWR Nebulizer MN 300 2 We grant you a 6 months warranty on the integrated battery cells of the control unit of the AWR Nebulizer MN 300 2 The AWR I...

Page 31: ......

Page 32: ...Are you happy with the AWR Nebulizer MN 300 2 Tell your friends about it we d be delighted If not Tell us and we will try to help ...