Requirements for transport and storage:

•  No corrosion gas and well-ventilated room.

•  Keep the device out of the reach of children.

•  Do not store the device in conditions such as direct sunlight, high temperature, humidity, dust or water infiltrations, etc.

•  Keep the device away from sloping surfaces, vibrations or shocks.

•  Transportation adopts general transportation means or follows the contract requirements. Avoid violent shock, vibration, rain 

and snow splash during the process of transportation.

5.4 Pollution-free disposal and recycling

The service life of product is 3 years. If the device exceeds the period of use, it must be discarded. Please contact the manufac-

turer or distributor for more information.

1) The atomizer out of use can be sent back to the manufacturer or distributor for proper recycling.

2) Used parts can be returned to the manufacturer or distributor for disposal, or treated in accordance with relevant laws and 

regulations.

Chapter 6 
TROUBLESHOOTING

17

ENGLISH

Problems

Reason analysis

Solutions

The device can't startup.

Battery is not well installed.

Check the installation of battery, and reinstall 

the batteries.

No atomizing or little atomizing 

appears when device working.

Medication cup is not well installed.

Check the installation of medication cup, and 

reinstall it. 

No medication in medication cup. 

Trickle  medication  into  medication  cup,  re-

member not to exceed its maximum capacity.

Improper medication.

Consult a doctor about whether the medica-

tion is suitable for the device.

The nebulizing sheet is dirty.  

Clean medication cup.

There is water around the nozzle 

of nebulizer. 

Due to temperature differences, the temper-

ature of medication cup surface is relatively 

low, medication mist comes in contact with the 

nozzle, then condenses into water droplets.

Remove medication cup, pour the water out.

After startup, power indicator 

lights about. 1s, then immediate-

ly goes out.

Medication cup is not well installed.

Install medication cup once again.

Medication cup is not loaded with any med-

ication.

Put the medication into medication cup after 

consulting your doctor.

After turning on the device, the 

power indicator lights once, then 

it is out immediately or the de-

vice cannot work normally.

The battery had run down.

Replace the batteries immediately.

Nebulizer,  doesn't  automatically 

shutdown when medication is 

used up.

Medication may generate bubbles in medica-

tion cup.

Press “ON/OFF” button to turn off the device, 

and clear up the bubbles.

Medication  may  be  stuck  on  the  nebulizing 

sheet.

Press “ON/OFF” button to turn off the device, 

and clean medication cup.

The electrodes contacting with the medication 

cup may be dirty. 

Press “ON/OFF” button to turn off the device, 

and clean the electrodes.

If the device still can't work normally alter using all methods above, please contact our after-sales service.

Summary of Contents for 28066

Page 1: ...TTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produ...

Page 2: ...thetic mixture Otherwise it may cause personal injury Do not wrap the power cord around children s neck or other sensitive areas or it may cause asphyxia The mask of this equipment is made of PVC mate...

Page 3: ...batteries is DC3V 2 3 Operational environment Temperature 5 C 40 C Humidity 15 90 Atmospheric pressure 700hPa 1060hPa Attention This product is not suitable for use in strong electromagnetic interfer...

Page 4: ...cup to the main unit be sure to install it properly until clasp sound is heard Otherwise it may cause electrode conduction failure then the device cannot atomize normally 2 Remove medication cup from...

Page 5: ...ery life will be affected Remove the batteries if the device won t be used for long time 4 2 Operations for treatment use Preparations before use 1 Remove medication cup please clean and disin fect it...

Page 6: ...the cup lid and discard residual medication 3 Add 75 ethanol solution in medication cup cover the cup lid then leave for at least 10mins it is advisable to gently shake it for better disinfection 4 Im...

Page 7: ...medication cup Trickle medication into medication cup re member not to exceed its maximum capacity Improper medication Consult a doctor about whether the medica tion is suitable for the device The ne...

Page 8: ...ode Keep away from sunlight 55 C 40 C Store between 40 and 55 C Lot number Fragile handle with care Date of manufacture Serial number Follow instructions for use Medical Device complies with Directive...

Page 9: ...4 2 Air Discharge 2 kV 4 kV 8 kV 15 kV contact 8 kV air Air Discharge 2 kV 4kV 8 kV 15 kV contact 8 kV air Floors should be wood concrete or ce ramic tile If floor is covered with synthetic material...

Page 10: ...ct was purchased If the equipment is not disposed of correctly fines or penalties may be applied in accordance with the national legislation and regulations GIMA WARRANTY CONDITIONS Congratulations fo...

Reviews: