4.  Install medication cup to main unit.

5.  Assemble the mask, as shown to the side:

Operation method:

1.  Turn on the power: Press "ON/OFF" button for more than 1 sec-

ond, power indicator (green) lights and device starts atomizing.

Attention:

If medication cup is not loaded with any medication, the de-

vice will automatically shutdown after power Indicator lights up 

about 1s.

After startup, the quantity of medication mist may change at the 

beginning of device working, which Is a normal phenomenon.

2.  Inhalation: Hold the device in hand, put on the mask, slowly 

inhale the medication mist.

Attention:

1) The angle of inclination should be within 45° during nebulizing.

2) During use, please do not strongly shake the device to prevent abnormal use.

3) Duration of each inhalation should be no more than 20 minutes.

4) Nebulizing treatment Is easy and comfortable, if you feel uncomfortable during use, please stop the treatment.

3.  Turn off the power. When the treatment is over, and medication almost runs out, the nebulizing sheet will generate a high-fre-

quency sound, and then the device automatically turns off. If you need to shut the device down during use, please press the 

"ON/OFF" button for more than 1 second.

Attention:

At the end of treatment, it is normal that a little medication remains left in medication cup after automatic shutdown.

Chapter 5 
MAINTENANCE, TRANSPORT AND STORAGE

5.1 Cleaning and disinfection

Clean and disinfect the device after each use. If the device is not cleaned, the nebulizing will be affected because of drying and 

coagulation of medication.

1.  Remove medication cup, accessory and batteries from the main unit.

2.  Open the cup lid and discard residual medication.

3.  Add 75% ethanol solution in medication cup, cover the cup lid then leave for at least 10mins; it is advisable to gently shake it 

for better disinfection.

4.  Immerse the accessories to be disinfected into a container with ethanol solution, and put a lid on the container. Use 75% 

ethanol solution soaking for 10mins or longer.

5.  Discard the disinfectant in medication cup, take accessories out from the disinfectant; clean the medication cup and accesso-

ries with clear-water repeatedly. 

6.  Fill medication cup with clear-water, assemble it to the main unit, let the device work 10mins in order to clean the nebulizing 

sheet.

7.  After cleaning, use new medical gauze to wipe away the water, and fully dry,

8.  Use 75% medicinal alcohol to wipe the surface of main unit, then air-dry or wipe-dry with a clean, soft cloth.

9.  After all steps above, store the main unit, medication cup and accessories in a dry, clean place.

Attention:

Do not throw medication cup and accessories into boiling water for disinfection, otherwise the part may be deformed. Do not 

put them in a microwave oven for drying.

The parts disinfected with disinfectant must be fully cleaned, or the residual disinfectant may cause symptomatic deterioration.

5.2 Medication cup replacement

The nebulizing sheet is a kind of consumable. In general, the service life of the atomizer is about six months (20 minutes per 

time, three times a day).

Its service life depends on the use, medication, and the degree of cleaning. If no atomizing or little atomizing appears when the 

device is working, please replace medication cup in time. (If you need to purchase medication cup, please contact the dealer.)

5.3 Transport and storage

Transport and storage environment:

Temperature: -40°C~+55°C

Relative humidity: 5%~96%

Atmospheric pressure: 500 hPa ~ 1060 hPa

ENGLISH

Summary of Contents for 28066

Page 1: ...TTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produ...

Page 2: ...thetic mixture Otherwise it may cause personal injury Do not wrap the power cord around children s neck or other sensitive areas or it may cause asphyxia The mask of this equipment is made of PVC mate...

Page 3: ...batteries is DC3V 2 3 Operational environment Temperature 5 C 40 C Humidity 15 90 Atmospheric pressure 700hPa 1060hPa Attention This product is not suitable for use in strong electromagnetic interfer...

Page 4: ...cup to the main unit be sure to install it properly until clasp sound is heard Otherwise it may cause electrode conduction failure then the device cannot atomize normally 2 Remove medication cup from...

Page 5: ...ery life will be affected Remove the batteries if the device won t be used for long time 4 2 Operations for treatment use Preparations before use 1 Remove medication cup please clean and disin fect it...

Page 6: ...the cup lid and discard residual medication 3 Add 75 ethanol solution in medication cup cover the cup lid then leave for at least 10mins it is advisable to gently shake it for better disinfection 4 Im...

Page 7: ...medication cup Trickle medication into medication cup re member not to exceed its maximum capacity Improper medication Consult a doctor about whether the medica tion is suitable for the device The ne...

Page 8: ...ode Keep away from sunlight 55 C 40 C Store between 40 and 55 C Lot number Fragile handle with care Date of manufacture Serial number Follow instructions for use Medical Device complies with Directive...

Page 9: ...4 2 Air Discharge 2 kV 4 kV 8 kV 15 kV contact 8 kV air Air Discharge 2 kV 4kV 8 kV 15 kV contact 8 kV air Floors should be wood concrete or ce ramic tile If floor is covered with synthetic material...

Page 10: ...ct was purchased If the equipment is not disposed of correctly fines or penalties may be applied in accordance with the national legislation and regulations GIMA WARRANTY CONDITIONS Congratulations fo...