Premium pipetting comfort!

EXTREMELY LOW PIPETTING FORCEs! 

EXTREMELY EAsY TIP EJECTION! 

LEGENdARY PIPETMAN RObusTNEss!

Ad

just

Able-Vol

ume

 Pi

Pet

te

Trusted results

Best results guaranteed with a performance exceeding ISO recommendations.

Model

Volume 

tested

Maximum Permissible Errors

Gilson

ISO 8655

Systematic error 

(μL)

Random error 

(μL)

Systematic error 

(μL)

Random error 

(μL)

P2G

(F144054P)
(F144054M)

Min.:  

0.2 

 0.5

Max.: 2

± 0.024

± 0.025

± 0.030

≤ 0.012

≤ 0.012

≤ 0.014

± 0.08

± 0.08

± 0.08

≤ 0.04

≤ 0.04

≤ 0.04

P10G

(F144055P)
(F144055M)

Min.:  

 5

Max.: 10

± 0.025

± 0.075

± 0.100

≤ 0.012

≤ 0.030

≤ 0.040

± 0.12

± 0.12

± 0.12

≤ 0.08

≤ 0.08

≤ 0.08

P20G

(F144056P)
(F144056M)

Min.:  

 10

Max.: 20

± 0.10

± 0.10

± 0.20

≤ 0.030

≤ 0.050

≤ 0.060

± 0.2

± 0.2

± 0.2

≤ 0.1

≤ 0.1

≤ 0.1

P100G

(F144057P)
(F144057M)

Min.:  

10 

 50

Max.: 100

± 0.35

± 0.40

± 0.80

≤ 0.10

≤ 0.12

≤ 0.15

± 0.8

± 0.8

± 0.8

≤ 0.3

≤ 0.3

≤ 0.3

P200G

 

(F144058P)
(F144058M)

Min.:  

20 

 100

Max.:  

200

± 0.50

± 0.80

± 1.60

≤ 0.20

≤ 0.25

≤ 0.30

± 1.6

± 1.6

± 1.6

≤ 0.6

≤ 0.6

≤ 0.6

P1000G

(F144059P)
(F144059M)

Min.:  

100 

 500

Max.: 1000

± 3.0

± 4.0

± 8.0

≤ 0.6

≤ 1.0

≤ 1.5

± 8.0

± 8.0

± 8.0

≤ 3.0

≤ 3.0

≤ 3.0

P5000G

(F144066)

Min.:  

500 

 2500

Max.: 5000

± 12

± 15

± 30

≤ 3

≤ 5

≤ 8

± 40

± 40

± 40

≤ 15

≤ 15

≤ 15

P10mL G

(F144067)

Min.:  

1000 

 5000

Max.: 10000

± 30

± 40

± 60

≤ 6

≤ 10

≤ 16

± 60

± 60

± 60

≤ 30

≤ 30

≤ 30

Gilson’s Complete Pipetting System

Your protocols and experiments require more than just generic pipette tips. Indeed, tips 
are essential components of your pipetting system. According to ISO 8655-2*:
“The maximum permissible errors always apply to the total system of piston pipette 
and tip. When using pipette tips not delivered by the pipette supplier, the supplier 
declaration or certificate of conformity does not apply”. 

*ISO 8655 is an international standard for “piston operated apparatus”. It defines, among other things, terms and 
definitions regarding piston operated pipettes and their testing, testing procedure and maximum permissible errors.

Pipette model

Recommended PIPETMAN DIAMOND tips

Model 

(scale 1/2)

Volume range

P2G

D10

0.1 – 10 µL

DL10

0.1 – 20 µL

DF10

0.1 – 10 µL

DFL10

0.1 – 10 µL

P10G

D10

0.1 – 10 µL

DL10

0.1 – 20 µL

DF10

0.1 – 10 µL

DFL10

0.1 – 10 µL

P20G

D200

2 – 200 µL

DF30

2 – 30 µL

P100G

D200

2 – 200 µL

DF100

20 – 100 µL

P200G

D200

2 – 200 µL

D300

20 – 300 µL

DF200

20 – 200 µL

DF300

20 – 300 µL

P1000G

D1000

100 – 1000 µL

D1200

100 – 1200 µL

DF1000

100 – 1000 µL

DF1200

100 – 1200 µL

P5000 G

D5000

1 – 5 mL

P10mL G

D10mL

1 – 10 mL

PIPETMAN G kits complement the product range and offer 
cost-effective package to get you started or to renew your 
old set of pipette. 

Pipette Kits

All you need to get started, in a single box!

PIPETMAN G

Ordering reference

Pipette Models*

PIPETMAN DIAMOND Tips

STARTER KIT 

F167900

P20G - P200G - P1000G

D200 - D1000

MICRO-VOLUME KIT

F167800

P2G - P10G - P100G

DFL10ST - DF100ST

*With both ejectors: stainless steel and plastic ejectors.

POS. 1 066803

POS. 2 066805

POS. 3 066807

POS. 4 066811

POS. 5 066813

Summary of Contents for Pipetman P1000G

Page 1: ...5000 12 15 30 3 5 8 40 40 40 15 15 15 P10mL G F144067 Min 1000 5000 Max 10000 30 40 60 6 10 16 60 60 60 30 30 30 Gilson s Complete Pipetting System Your protocols and experiments require more than ju...

Page 2: ...for the most intensive pipetting routine Thanks to the redesigned piston assembly new seal lighter springs and the addition of high quality pis ton grease you can enjoy one of the lowest pipetting fo...

Page 3: ...lson ISO 8655 Errore Errore Errore Errore Sistematico Casuale Sistematico Casuale P2G F144054P F144054M Min 0 2 0 5 Max 2 0 024 0 012 0 025 0 012 0 030 0 014 0 08 0 04 0 08 0 04 0 08 0 04 P10G F144055...

Page 4: ...RTANT NOTICE This pipette can be only lubricated with a high quality lubricant selected and approved by Gilson This lubricant is availableunderthereference 5440011070 Theuseofothertypeoflubricantcance...

Page 5: ...plasticorastainless steel ejector Implementation of new features to improve dramatically pipettingcomforthasnoimpactonthelegendaryrobustness accuracy and precision of Pipetman Eight single channel mod...

Page 6: ...son distributor 2 Parts Check List Just take a moment to verify that the following items are present Pipetman G User s Guide Safety bag Certi cate of conformity including bar code sticker 3 Descriptio...

Page 7: ...ing slowlyreachthe requiredsetting makingsurenottoovershootthemark whenincreasingthevolumesetting passtherequired valueby1 3ofaturnandthenslowlydecreasetoreach the volume making sure not to overshoot...

Page 8: ...ed with the adapter in place ready to use DL10 tips If D10 tips are used the adapter must be repositioned in the shorter slot as follows Pull the adapter down from the metallic rod Turn the adapter th...

Page 9: ...e levels of accuracy and precision within specifications Aspirate Press the push button to the rst stop this corresponds to the set volume of liquid Holdthepipetteverticallyandimmersethetipinthe liqui...

Page 10: ...thly When aspirating keep the tip at a constant depth below the surface of the liquid refer to the table Change the tip before as pirating a different liquid sample or reagent Change the tip if a drop...

Page 11: ...Chapter 7 Accessories Do not pipette liquids having temperatures above 70 Corbelow4 C Thepipettecanbeusedbetween 4 C and 40 C but the speci cations may vary accordingtothetemperature referto theISO86...

Page 12: ...rmity provide traceability of your pipette 9 Troubleshooting Aquickinspectionofthepipettemayhelpyoutodetect a problem YoumaydownloadfromtheGilsonwebsite www gilson com the 2minuteinspection whichshows...

Page 13: ...ting nut Pipette is Tip holder is loose Tighten connecting nut not precise Connecting nut is loose Tighten connecting nut Incorrect operator technique Operator training Damaged or corroded piston s Re...

Page 14: ...om a beaker of distilled water Maintainthepipetteintheverticalpositionandwait for 20 seconds If a water droplet appears at the end of the tip there is a leak If you see no droplet re immerse the tip b...

Page 15: ...r and its adapter With these pipettes if the tip holder is damaged the piston may also be damaged After replacing any parts you should verify the performance of your pipette following the veri cation...

Page 16: ...13 P10G P20G P100G P200G P5000G P10mlG P1000G Tip holder Tip ejector stainless steel plastic Connecting nut Color coded push button Piston assembly Blue ring...

Page 17: ...g pur posesonly Ifthepistonassemblyischanged thepipette must be adjusted and calibrated in a Gilson authorized Service Center As the models P2G and P10G contain miniaturized parts it is best not to di...

Page 18: ...del of pipette Remove the tip ejector see above Unscrew the connecting nut by turning it counter clockwise Carefully separate the lower and upper parts Remove the piston assembly O ring and seal If ne...

Page 19: ...nt free tissue soaked with distilled water Re t the tip ejector and allow the pipette to dry Internal The following components only can be immersed in a cleaning solution connecting nut tip ejector ti...

Page 20: ...You may choose to decontaminate your pipette chemi cally inaccordancewithyourownprocedures Whatever decontaminant you use check with the supplier of the decontaminant that it is compatible with stainl...

Page 21: ...is a high quality pipette that offers excel lent accuracy and precision The gures given in the Gilson Maximum Permissible Errors table page 19 were obtained using PIPETMAN Diamond Tips These gures ar...

Page 22: ...20 0 50 0 20 1 60 0 60 F144058P 100 0 80 0 25 1 60 0 60 F144058M Max 200 1 60 0 30 1 60 0 60 P1000G Min 100 3 0 6 8 3 0 F144059P 500 4 1 0 8 3 0 F144059M Max 1000 8 1 5 8 3 0 P5000G Min 500 12 3 40 1...

Page 23: ...ngs 1 tip holder ServiceKit2nd levelincludes 1 push button 1 connecting nut 2 tip ejectors for the P2G and P10G models only 1 adapter for stainless steel tip ejector 1 extension for plastic tip ejecto...

Page 24: ...tor adapter F144879 F144879 F4 Tip ejector extension F2070903 F2070903 P5000G F144066 and P10mlG F144067 Description P5000G P10mlG C D E Service Kit 1st level FA07021 FA07022 A Push button assembly FA...

Page 25: ...of the following European Directives Villiers le Bel May 3rd 2012 A El Sayed H Le Dorze General Manager Quality Manager GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France Te...

Page 26: ...shall also be void in theeventpipettesarealteredormodifiedbyanypartyotherthan Gilson or its designates Gilson s sole liability under this warranty shallbelimitedto atGilson ssoleoption repairorreplace...

Page 27: ...lson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 I Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I BP...

Page 28: ...RTANT NOTICE This pipette can be only lubricated with a high quality lubricant selected and approved by Gilson This lubricant is availableunderthereference 5440011070 Theuseofothertypeoflubricantcance...

Page 29: ...nejectorplasticorastainless steel ejector Implementation of new features to improve dramatically pipettingcomforthasnoimpactonthelegendaryrobustness accuracy and precision of Pipetman Eight single cha...

Page 30: ...son distributor 2 Parts Check List Just take a moment to verify that the following items are present Pipetman G User s Guide Safety bag Certi cate of conformity including bar code sticker 3 Descriptio...

Page 31: ...ing slowlyreachthe requiredsetting makingsurenottoovershootthemark whenincreasingthevolumesetting passtherequired valueby1 3ofaturnandthenslowlydecreasetoreach the volume making sure not to overshoot...

Page 32: ...ed with the adapter in place ready to use DL10 tips If D10 tips are used the adapter must be repositioned in the shorter slot as follows Pull the adapter down from the metallic rod Turn the adapter th...

Page 33: ...e levels of accuracy and precision within specifications Aspirate Press the push button to the rst stop this corresponds to the set volume of liquid Holdthepipetteverticallyandimmersethetipinthe liqui...

Page 34: ...thly When aspirating keep the tip at a constant depth below the surface of the liquid refer to the table Change the tip before as pirating a different liquid sample or reagent Change the tip if a drop...

Page 35: ...Chapter 7 Accessories Do not pipette liquids having temperatures above 70 Corbelow4 C Thepipettecanbeusedbetween 4 C and 40 C but the speci cations may vary accordingtothetemperature referto theISO86...

Page 36: ...rmity provide traceability of your pipette 9 Troubleshooting Aquickinspectionofthepipettemayhelpyoutodetect a problem YoumaydownloadfromtheGilsonwebsite www gilson com the 2minuteinspection whichshows...

Page 37: ...ting nut Pipette is Tip holder is loose Tighten connecting nut not precise Connecting nut is loose Tighten connecting nut Incorrect operator technique Operator training Damaged or corroded piston s Re...

Page 38: ...om a beaker of distilled water Maintainthepipetteintheverticalpositionandwait for 20 seconds If a water droplet appears at the end of the tip there is a leak If you see no droplet re immerse the tip b...

Page 39: ...r and its adapter With these pipettes if the tip holder is damaged the piston may also be damaged After replacing any parts you should verify the performance of your pipette following the veri cation...

Page 40: ...13 P10G P20G P100G P200G P5000G P10mlG P1000G Tip holder Tip ejector stainless steel plastic Connecting nut Color coded push button Piston assembly Blue ring...

Page 41: ...pur posesonly Ifthepistonassemblyischanged thepipette must be adjusted and calibrated in a Gilson authorized Service Center As the models P2G and P10G contain miniaturized parts it is best not to dis...

Page 42: ...del of pipette Remove the tip ejector see above Unscrew the connecting nut by turning it counter clockwise Carefully separate the lower and upper parts Remove the piston assembly O ring and seal If ne...

Page 43: ...free tissue soaked with distilled water Re t the tip ejector and allow the pipette to dry Internal The following components only can be immersed in a cleaning solution connecting nut tip ejector tip h...

Page 44: ...You may choose to decontaminate your pipette chemi cally inaccordancewithyourownprocedures Whatever decontaminant you use check with the supplier of the decontaminant that it is compatible with stainl...

Page 45: ...gh quality pipette that offers excel lent accuracy and precision The gures given in the Gilson Maximum Permissible Errors table page 19 were obtained using PIPETMAN Diamond Tips These gures are guaran...

Page 46: ...20 0 50 0 20 1 60 0 60 F144058P 100 0 80 0 25 1 60 0 60 F144058M Max 200 1 60 0 30 1 60 0 60 P1000G Min 100 3 0 6 8 3 0 F144059P 500 4 1 0 8 3 0 F144059M Max 1000 8 1 5 8 3 0 P5000G Min 500 12 3 40 1...

Page 47: ...ngs 1 tip holder ServiceKit2nd levelincludes 1 push button 1 connecting nut 2 tip ejectors for the P2G and P10G models only 1 adapter for stainless steel tip ejector 1 extension for plastic tip ejecto...

Page 48: ...tor adapter F144879 F144879 F4 Tip ejector extension F2070903 F2070903 P5000G F144066 and P10mlG F144067 Description P5000G P10mlG C D E Service Kit 1st level FA07021 FA07022 A Push button assembly FA...

Page 49: ...of the following European Directives Villiers le Bel March 1st 2011 A El Sayed H Ledorze General Manager Quality Manager GILSON S A S 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel France T...

Page 50: ...NOTES...

Page 51: ...ilson Inc World Headquarters 3000 Parmenter Street I P O Box 620027 I Middleton WI 53562 0027 USA Tel 1 800 445 7661 or 1 608 836 1551 I Fax 1 608 831 4451 Gilson S A S 19 avenue des Entrepreneurs I B...