Upper Part (handle)
Wipe the upper part (handle) of the pipette with
a soft-cloth or lint-free tissue impregnated with
the chosen decontaminant.
Wipe the upper part of the pipette with a soft-
cloth or lint-free tissue soaked with distilled water
or sterile water.
Lower Part (Volumetric module)
The following components
only
can be immersed in a
decontaminant solution: connecting nut, tip-ejector,
tip-holder, piston assembly, seal and o-ring.
13 - sPecIFIcatIons
PIPETMAN P is a high quality pipette that offers excel-
lent accuracy and precision. The figures given in the
“Gilson Maximum Permissible Errors” table (page 19)
were obtained using Gilson DIAMoND Tips. These
figures are guaranteed only when genuine Gilson
DIAMoND Tips are used.
Each pipette is inspected and validated by qualified
technicians in accordance with the Gilson Quality
System. Gilson declares that its manufactured
pipettes comply with the requirements of the ISo
8655 standard, by type testing.
The adjustment is carried out under strictly defined
and monitored conditions (ISo 8655-6).