19
Gilson maximum Permissible errors
Single channel
Maximum Permissible Errors
Model
Volume
Gilson
ISO 8655
(Reference)
(µL)
systematic
Random
systematic
Random
error (µL)
error (µL)
error (µL)
error (µL)
Min
0.2 ± 0.024
0.012
± 0.08
0.04
P2
0.5 ± 0.025
0.012
± 0.08
0.04
(f144801)
Max.
2 ± 0.030
0.014
± 0.08
0.04
Min.
1 ± 0.025
0.012
± 0.12
0.08
P10
5 ± 0.075
0.030
± 0.12
0.08
(f144802)
Max
10 ± 0.100
0.040
± 0.12
0.08
Min.
2 ± 0.10
0.03
± 0.20
0.10
P20
5 ± 0.10
0.04
± 0.20
0.10
(f123600)
10 ± 0.10
0.05
± 0.20
0.10
Max.
20 ± 0.20
0.06
± 0.20
0.10
Min.
20 ± 0.35
0.10
± 0.80
0.30
P100
50 ± 0.40
0.12
± 0.80
0.30
(f123615)
Max.
100 ± 0.80
0.15
± 0.80
0.30
Min.
50 ± 0.50
0.20
± 1.60
0.60
P200
100 ± 0.80
0.25
± 1.60
0.60
(f123601)
Max.
200 ± 1.60
0.30
± 1.60
0.60
Min.
200 ± 3
0.6
± 8
3.0
P1000
500 ± 4
1.0
± 8
3.0
(f1123602) Max.
1000 ± 8
1.5
± 8
3.0
Min.
1000 ± 12
3
± 40
15
P5000
2000 ± 12
5
± 40
15
(f123603)
Max.
5000 ± 30
8
± 40
15
Min.
1 mL ± 30
6
± 60
30
P10ml
2 mL ± 30
6
± 60
30
(f161201)
5 mL ± 40
10
± 60
30
Max.
10 mL ± 60
16
± 60
30
With a precise pipetting technique (see Chapter 6 - General guide-
lines for good pipetting) the P2 model may be used to aspirate
volumes as low as 0.1 µL and the P10 model as low as 0.5 µL.
The data given in the tables conform to the ISO 8655-2 Standard.