MAXIMUM PERMISSIBLE ERRORS
MODEL
(PART
NUMBER)
VOLUME
(µL)
GILSON
ISO 8655
SYSTEMATIC
ERROR
(µL)
RANDOM
ERROR
(µL)
SYSTEMATIC
ERROR
(µL)
RANDOM
ERROR
(µL)
M10E
(FD10001)
Min.
1
±0.09
≤
0.03
±0.2
≤
0.1
5
±0.10
≤
0.03
±0.2
≤
0.1
Max.
10
±0.15
≤
0.06
±0.2
≤
0.1
M25E
(FD10002)
Min.
3
±0.25
≤
0.08
±0.7
≤
0.3
10
±0.27
≤
0.08
±0.7
≤
0.3
Max.
25
±0.30
≤
0.10
±0.7
≤
0.3
M50E
(FD10003)
Min.
20
±0.34
≤
0.20
±0.7
≤
0.3
Max.
50
±0.70
≤
0.30
±0.7
≤
0.3
M100E
(FD10004)
Min.
10
±0.50
≤
0.20
±1.5
≤
0.6
50
±0.75
≤
0.30
±1.5
≤
0.6
Max.
100
±1.00
≤
0.40
±1.5
≤
0.6
M250E
(FD10005)
Min.
50
±1.50
≤
0.30
±6
≤
2.0
100
±1.70
≤
0.30
±6
≤
2.0
Max.
250
±2.50
≤
0.50
±6
≤
2.0
M1000E
(FD10006)
Min.
100
±3.0
≤
1.6
±12
≤
4.0
500
±5.0
≤
2.5
±12
≤
4.0
Max.
1000
±8.0
≤
4.0
±12
≤
4.0
FI
T
TI
N
G
T
h
E
C
A
p
IL
L
A
R
Y
p
IS
TO
N
8
MICROMAN E®
|
USER’S GUIDE
INTRODUCTION
|
MICROMAN E®
Chapter 2
FITTING ThE CApILLARY
pISTON
Make sure the capillary piston
(CP) is compatible with the
MICROMAN E model that will
be used.
CP25, CP50, and CP250 Bulk
The capillaries and pistons
are delivered in two separate
boxes. To fit a capillary piston :
1.
Take care not to
damage the sealing
tip when handling
the piston .
2. Select a piston
and slide it into the
capillary .
3. Gently push the
capillary until it snaps
onto the capillary
holder.
4. Attach the stem (
by slowly pressing
the push button until
you feel and hear a
slight click and then
continue to press to
the first stop. Finally,
pipette the liquid.
CP
MODEL
CP10
M10E
CP25
M25E
CP50
M50E
CP100
M100E
CP250
M250E
CP1000 M1000E
1
3
3
4
2
1
4
3
2
Figure 3
MICROMAN
Capillary Piston
Never lubricate the capillary holder,
capillary, or any other part of the
pipette. If the capillary tends to slip
off the capillary holder, clean the
capillary holder with ethanol, using
a medical wipe or similar soft tissue.
NOTICE
Summary of Contents for microman e