INTRO
D
U
C
TIO
N
6
MICROMAN E®
|
USER’S GUIDE
INTRODUCTION
|
MICROMAN E®
Specifications
MICROMAN E is a high quality pipette that offers
excellent accuracy and precision. The figures
provided in the “ Gilson Maximum Permissible
Errors ” table were obtained using Gilson CPs. These
figures are only guaranteed when using genuine
Gilson CPs.
Each pipette is inspected and validated by qualified
technicians according to the Gilson Quality System.
Gilson declares that its manufactured pipettes
comply with the requirements of the ISO 8655
Standard, by type testing. The adjustment is carried
out under strictly defined and monitored conditions
(ISO 8655-6):
●
Basis of adjustment, Ex.
●
Reference temperature, 20°C
●
Relative humidity, 50%
●
Barometric pressure, 101 kPa
●
Use of distilled water grade 3 (ISO 3696)
●
Ten measurements for each test volume,
which are nominal volume, 50% of nominal
volume, and the minimum or 10% of nominal
volume.
Performance Tests
Each pipette is inspected and validated according
to the Gilson Quality Assurance System. Based
on extensive historical data, manufacturing
conditions, and expertise, and in compliance with
ISO standards relative to statistical process control,
the assurance level of this instrument performing
to specifications is 99.8 %. Specifications rely on
quality and consistency for the whole pipetting
system and are guaranteed only when the pipette is
used with Gilson capillary pistons.
Summary of Contents for microman e
Page 1: ...User s Guide EN E...