14
MICROMAN E®
|
USER’S GUIDE
TROUBLESHOOTING
|
MICROMAN E®
Chapter 5
CLEANING AND
DECONTAMINATION
MICROMAN E is designed so that parts normally
in contact with liquid contaminants are easy to
clean and decontaminate.
Cleaning
Clean the pipette with soap solution prior to
decontamination.
External
1.
Wipe the entire pipette with a soft cloth or
lint-free tissue soaked with soap solution
to remove any stains. If the pipette is very
dirty, a brush with soft plastic bristles may
be used.
2. Wipe the entire pipette with a soft cloth or
lint-free tissue soaked with distilled water.
3. Leave the parts to dry by evaporation or
wipe them down with a clean soft cloth or
lint-free tissue.
Internal
Only the following components should be
immersed in a decontaminant solution: clamp
assembly, return spring, and capillary holder
(shaft).
1.
Unscrew the capillary holder (shaft).
2. Remove the clamp assembly and return
spring from the shaft.
3. Set aside the body (handle) in a dry and
secure location.
WARNING
Before returning any pipette,
ensure that it is completely
free of chemical, biological, or
radioactive contamination. Use
the safety bag provided by Gilson.