6
CENTRY™ 117 MiCRoCENTRifugE
| uSER’S guiDE
iNSTALLATioN
|
CENTRY™ 117 MiCRoCENTRifugE
Installation
Unpacking
Before unpacking the CENTRY 117 Microcentrifuge, inspect
the outside of the carton for any shipping damage. The
CENTRY 117 Microcentrifuge is delivered in a carton with
protective foam cushioning.
1. Remove the CENTRY 117 Microcentrifuge from the carton.
2. Retain the carton and cushioning until it has been
established that the CENTRY 117 Microcentrifuge is
working.
NOTE
Shipping damage is the responsibility of the transportation
carrier. Any claims for damage must be filed within 48 hours
with the carrier.
The accessories and instruction manual supplied with the
CENTRY 117 Microcentrifuge should be kept with the near its
place of installation.
Installation Site Requirements
The CENTRY 117 Microcentrifuge should be installed on a rigid,
even surface such as a stable laboratory bench, countertop,
etc. To guarantee sufficient ventilation, ensure that the
microcentrifuge has at least 15 cm (6 inches) of free space on
all sides, including the rear.
NOTICE
The CENTRY 117 Microcentrifuge should not be located in
areas subject to excessive heat such as in direct sunlight or
near radiators or the exhaust of a compressor, as a buildup of
heat may occur within the chamber.