12
CENTRY™ 117 MiCRoCENTRifugE
| uSER’S guiDE
MAiNTENANCE
|
CENTRY™ 117 MiCRoCENTRifugE
Replace Fuses
Check the fuse when recommended (refer to
1. Disconnect the power cord from the power inlet.
2. Locate the fuse holder in the power inlet on the rear panel.
3. open the fuse holder drawer by inserting a small
screwdriver under the tab and prying it open.
4. Remove the innermost (operative) fuse from its retaining
tabs and replace the fuse if necessary.
NOTE
A spare fuse is located in the outermost chamber of the
fuse drawer.
Replace only with a fuse of exactly the same value as the
original.
Fuse type may be found in the Specifications section of
this manual.