CENTRY™ 117 MiCRoCENTRifugE
| uSER’S guiDE
MAiNTENANCE
|
CENTRY™ 117 MiCRoCENTRifugE
Troubleshooting
Overloading Rotors
The maximum load of the rotor and the maximum speed have
been established by the manufacturer.
CAUTION
Do not attempt to exceed the maximum load and
maximum speed as established by the manufacturer.
The maximum speed of the rotor has been established for
liquids having a homogeneous density of 1.2 g/mL or less. in
order to centrifuge liquids with a higher density it is necessary
to reduce the speed. failure to reduce the speed may result
in damage to the rotor and microcentrifuge. The revised
maximum speed can be calculated with the following formula:
SYMPToM
PoSSiBLE CAuSE
SoLuTioN
CENTRY 117
Microcentrifuge will
not start
No power supply
Check that power is
being supplied to the
outlet
Check that the power
cord is plugged into
both the wall outlet
and the back of the
centrifuge
Check that power cord
is not damaged
Blown fuse
Check fuse and replace
if necessary
TROUBLESHOOTING (CONTINUED ON PAGE 14)