Annex C.2 Acoustic Output Reporting Table
- 51 -
Annex C.2 Acoustic Output Reporting Table
Acoustic Output Reporting Table for Track1
Non-Autoscanning Mode
System: GRU-5000A Ultrasonic Biometer for Ophthalmology
Transducer Model: Prb1000 A/10-C
Operating
Mode:
A-Mode
Application:
Ophthalmology
Working
Frequency:
10MHz
Acoustic Output
MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Maximum index value
0.079 0.003 1.68
Associated
acoustic
parameter
P
r.3
(MPa)
0.241
W
0
(Mw)
0.00035
0.00035
f
c
(MHz)
9.203
9.203
9.203
Z
SP
(cm)
2.10
2.10
2.10
Beam dimensions
x
-6
(cm)
0.169
0.169
y
-6
(cm)
0.162
0.162
PD (usec)
0.202
0.202
PRF (Hz)
10
EDB
Az (cm)
Ф
0.48
Ele (cm)
Ф
0.48
Operating
Control
Conditions
--
--
--
--
--
--
--
--
Transducer Model:
Prb1000A/10-C
Operation Mode:
A Mode
Application:
Ophthalmology
Working Frequency:
10MHz
Acoustic Output
MI
I
SPAT.3
( mW/cm
2
)
I
SPPA.3
( W/cm
2
)
Sample Size
3 3 3
K ( P=0.90, gamma=0.9 )
4.258 4
.
258 4
.
258
Mean ( x )
0
.
071 0
.
003 1.613
StdDev ( Sx )
0
.
007 0
.
000 0.335
Limit ( X )
0.100 0.005 3.040