8003 Precision Scalar Analyzer
3-22
Manual 20791, Rev. C, June 2001
display, apply correct Cal Factors to the reading, take
the correct portions of trace memories and path cal
memories, and fix cursors and markers. The update
modifier allows independent control of the sweeper
frequencies and all other frequency related analyzer
functions. If the modifier UPD ON is included once with
the SWPF, SWPS, or FIXED verbs, then all subsequent
uses of those verbs will update the sweeper frequency
along with the analyzer frequency. If the modifier UPD
OFF is included once with the above verbs, then all
subsequent uses of those verbs will change only
analyzer frequency parameters.
Modifier name:
UPD
Modifier syntax:
UPD
hs
state
State definition:
ON, OFF
Send the following commands to modify both the sweeper and analyzer frequencies to
2000 MHz
FIXED;UPD ON
FIXED;2000
Send the following commands to modify only the analyzer frequency to 6000 MHz.
FIXED;UPD OFF
FIXED;6000
Description:
This verb places the signal source mode to fixed frequency. This is an instrument wide
command and, as such, has no channel or sensor specifiers. However there are many
parameters associated with the source mode. These are defined by the modifiers which
follow the verb.
Summary of Contents for 8003
Page 3: ......
Page 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Page 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Page 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Page 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Page 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Page 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Page 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Page 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Page 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...