Performance Test & Calibration
Manual 20791, Rev. C, June 2001
4-17
NOTES:
1.
Linearity Error (%) = [(R1/R2) / (P1/P2) - 1] x 100
2.
Accumulated error is the sum of the current dB segment linearity error plus the previous accumulated error.
3.
Use the first CW Linearity error value entered in the +16 dBm to +20 dBm Linearity Data on page 4-16.
Swept Linearity Data - (+16 dBm to -50 dBm)
Step
Attenuator
Value
Power
Set Point
Power
Meter
Reading
(P)
8540
(DUT)
Reading
(R)
Reference
Power
Ratio
8540
(DUT)
Reading
Ratio
Linearity Error (%)
1
Linearity
Specification
Accumulated
Linearity
Error
2
0 dB
See Note 3
±
1.4%
Same as Lin.
error above
10 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
1.4%
20 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
1.4%
30 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
1.4%
40 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
1.5%
50 dB
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
2.4%
60 dB
(Increase avg.
to 8)
1.00 mW
±
0.025 mW
P1 =
R1 =
P1/P2 =
R1/R2 =
10.00 mW
±
0.25 mW
P2 =
R2 =
±
11.4%
Summary of Contents for 8003
Page 3: ......
Page 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Page 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Page 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Page 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Page 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Page 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Page 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Page 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Page 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...