Remote Operation
Manual 20791, Rev. C, June 2001
3-19
Function
Default Setting
Source
Signal Source Mode:
Start/Stop Swept Freq
Start Frequency:
Sweeper Start Frequency or 1 MHz if not
controlled by the 8003
Stop Frequency:
Sweeper Stop Frequency
Sweep Time:
200 ms
RF Power Level:
OFF, minimum RF level of the signal source or
0 dBm if not controlled by the 8003.
Source Modulation:
OFF
The following instrument settings are
not
changed during an instrument preset:
•
All GPIB addresses (Analyzer, Source, Printers, and Plotters)
•
Sensor linearity calibrations
•
Trace Memories
•
Path Cal Memories
3.3.10
Disable Private Bus Command Structure
DISABLE
Syntax:
DISABLE
Related Command:
ENABLE
Description:
This verb disables the private bus. When sent, the 8003 private bus becomes inactive. An
Enable command must be sent or a power down/power up cycle must be executed to
enable the private bus.
3.3.11
Enable Private Bus Command Structure
ENABLE
Syntax:
ENABLE
Related Command:
DISABLE
Description:
This verb enables the private bus. When sent, the 8003 private bus becomes active. A
Disable command must be sent to disable the private bus.
Summary of Contents for 8003
Page 3: ......
Page 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Page 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Page 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Page 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Page 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Page 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Page 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Page 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Page 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...