Example programs for Channel Configuration are contained in Appendix A.
CALC#:UNIT
Syntax:
CALCulate<channel 1 to 2>:UNIT[:POWer]
space
<data units selection, dBm or Watt>
Example:
OUTPUT @PWR_MTR;CALC1:UNIT W
! SELECTS CHANNEL 1 LINEAR UNITS, WATTS
Example 2:
OUTPUT @PWR_MTR;CALC1:UNIT DBM
! SELECTS CHANNEL 1 LOG UNITS, DBM
Description:
This command configures a channel to report power measurements in either linear Watts
units or logarithmic dBm units. When UNIT is dBm, the RATio configuration produces
measurements in dB, and POWer and DIFF measurements in dBm. When UNIT is Watts, the
RATio configuration measurements are percentages, and POWer and DIFF measurements
are in Watts.
CALC#:STAT
Syntax:
CALCulate<channel 1 to 2>:STATe
space
<ON | OFF>
Example:
OUTPUT @PWR_MTR;CALC1:STAT OFF ! DISABLES CHANNEL 1 MEASUREMENT
Description:
This command turns a channel off or on. Default is ON. During Normal and Swift modes,
measurement speed increases when only one sensor (Channel Configuration is
CALC#:POW) is performing measurements.
CALC#:STAT?
Syntax:
CALCulate<channel 1 to 2>:STATe?
Example:
OUTPUT @PWR_MTR;CALC1:STAT?
! QUERIES CHANNEL 1 MEASUREMENT STATUS
Response:
0 = off, 1 = on
Description:
This command is the query format of the command above.
Summary of Contents for 58542
Page 3: ...DECLARATION OF CONFORMITY ...
Page 4: ......
Page 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Page 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Page 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Page 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Page 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Page 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21360 Rev A has not been u ...
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...