Calibration & Testing
Publication 21555, Rev. E, September 2002
4-11
4.4.2.1
Setup Parameters
The following setup parameters should be accomplished prior to performing the Power Linearity test:
4.4.2.2
Equipment Required
GPIB Controller
·
Digital Voltmeter
·
RF Signal Generator
·
Thermistor Power Meter
·
Directional Coupler
·
Step
Attenuator
·
Power Sensor
1.
The power meter and sensor should be calibrated by following the instructions in Section 4.3 of this
publication.
2.
The Averaging is set to AUTO by entering:
SENS1:AVER:COUN:AUTO ON
(Enter SENS2 for channel 2)
Extreme care is required in the following procedure since the ac-
curacy requirements are critical to ensure the most accurate test
results.
Figure 4-3: Power Sensor Linearity Test Setup
CAUTION
Summary of Contents for 58542
Page 3: ...DECLARATION OF CONFORMITY ...
Page 4: ......
Page 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Page 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Page 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Page 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Page 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Page 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21360 Rev A has not been u ...
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...