58542 VXIbus Universal Power Meter
2-2
Publication 21555, Rev. E, September 2002
2.3
Initial Setup
The logical address and data transfer bus arbitration must be set up in the power meter before installing
the unit in a VXI mainframe and applying power. The following procedures define how to complete the
initial setup.
2.3.1
Logical Address
The VXI chassis Resource Manager identifies each unit in the system by its logical address. The VXI
logical address can range from 0 to 255. Addresses 0 and 255 are reserved for special functions:
Address 0 identifies the Resource Manager (slot and controller); address 255 permits the Resource
Manager to dynamically address the unit based on the chassis VXI slot.
To change the logical address, set the respective sections on the eight position DIP switch. The switch is
accessible after removing the right cover (See Figure 2-1).
The address is set with binary values of 0 to 255. Switch position 1 is the least significant bit of the
address. Figure 2-1 illustrates logical address values of 3 (binary 00000011) and 255 (binary 11111111).
Giga-tronics ships the power meter with a logical address of 255 for dynamic configuration.
Figure 2-1: Setting the Logical Address
Summary of Contents for 58542
Page 3: ...DECLARATION OF CONFORMITY ...
Page 4: ......
Page 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Page 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Page 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Page 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Page 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Page 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21360 Rev A has not been u ...
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...