Introduction
Publication 21555, Rev. E, September 2002
1-5
1.2.3
Zeroing Accuracy
(Standard CW Sensors)
Notes:
1.
Depending on sensor used.
2.
Specified performance applies with maximum averaging and 24-hour warm-up at constant temperature.
System Linearity
(@ 50 MHz for Standard CW Sensors)
±0.02 dB over any 20 dB range from -70
-
+16 dBm
±0.02 dB + (0 dB, -0.05 dB/dB) from +16
-
+20 dBm
±0.04 dB from -70
-
+16 dBm
Linearity Temperature Coefficient
<0.1%/ °C temperature change following Power Sweep Calibration, 24-hour warm-up required
<0.3%/ °C temperature change following Power Sweep Calibration, 24-hour warm-up required (8035XA Series Sensors)
Zero Set
<
±
50 pW
2
Zero Drift
<
±
100 pW
during 1 hour
2
Noise
<
±
50 pW
measured over any 1 minute interval
2
Averaging
Auto-averaging or user-selectable averaging from 1 to 512 readings per measurement
Summary of Contents for 58542
Page 3: ...DECLARATION OF CONFORMITY ...
Page 4: ......
Page 18: ...58542 VXIbus Universal Power Meter xiv Publication 21555 Rev E September 2002 ...
Page 20: ...58542 VXIbus Universal Power Meter xvi Publication 21555 Rev E September 2002 ...
Page 30: ...58452 VXIbus Universal Power Meter 1 10 Publication 21555 Rev E September 2002 ...
Page 92: ...58542 VXIbus Universal Power Meter 2 62 Publication 21555 Rev E September 2002 ...
Page 104: ...58452 VXIbus Universal Power Meter 3 12 Publication 21555 Rev E September 2002 ...
Page 126: ...58452 VXIbus Universal Power Meter 5 8 Publication 21555 Rev E September 2002 ...
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21360 Rev A has not been u ...
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 248: ...58542 VXIbus Universal Power Meter C 2 Publication 21555 Rev E September 2002 ...