Performance Verification
2500A Series Operation Manual, 34172 Revision C, March 2008
3. Press the FM button to activate the FM Menu. If FM Menu 2 is not displayed, press the FM X/3
softkey until FM Menu 2 is displayed.
4. Press the FM softkey and the Step Up button to activate the Internal FM Modulation Generator. The
FM indicator will appear in the upper left corner of the display.
5. Press the Rate softkey and enter the first value on Datasheet 12. Measure and record the reading of
the frequency counter
6. Repeat Step 5 for the remaining data points on the datasheet.
4.3.3.4.3 Pulse Repetition Interval Accuracy
1. Connect the test equipment and UUT as shown in Figure 4-10. Press the Display Time button on the
Frequency Counter. Connect a BNC cable from the PM Out connector to the Band 1, BNC connec-
tor of the frequency counter
2. Set the CW frequency of the 2500A to 6 GHz.
3. Configure the counter to measure the period of the video pulse waveform.
4. Press the PM button to activate the PM Menu. If PM Menu 2 is not displayed, press the PM X/4
softkey until PM Menu 2 is displayed.
5. Press the PM softkey and the Step Up button to activate the Internal PM Modulation Generator. The
PM indicator will appear in the upper left corner of the display.
6. Press the PRI softkey and enter the first value on Datasheet 13.
7. Press the Width softkey and enter the first pulse width value on Datasheet 13. Measure and record
the reading of the frequency counter.
8. Repeat Step 5 for the remaining data points on the datasheet.
Summary of Contents for 2508A
Page 1: ...2500A Series 2500A Series Microwave Synthesizer Operations Manual 34172 Revision C March 2008 ...
Page 4: ......
Page 22: ...Chapter 1 2500A Introduction 22 2500A Series Operation Manual 34172 Revision C March 2008 ...
Page 88: ...2500A Chapter 3 Remote Operation 88 2500A Series Operation Manual 34172 Revision C March 2008 ...