Serial protocol
Continuous transmission protocol
This protocol is used for continuous transmission, usually towards the repeater panel.
COMMUNICATION PARAMETERS ARE SELECTABLE (see serial menu on page 39)
STRING TRANSMITTED AT A FREQUENCY OF 10 Hz:
STX <state> <net> ETX <csum> EOT
<state>
= character coded as in the following table (bit = 1 if condition TRUE)
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
0
0
1
1
Tare in-
sertet
Minimum
weighing
Weight
stable
Center
zero
<netto>
= fields consisting of 8 ASCII characters with the weight value justified on the right.
In overweight conditions, the field assumes the value: “^^^^^^^^”.
In underweight conditions, the field assumes the value: “_ _ _ _ _ _ _ _”.
Under weight reading error conditions, the field assumes the value: “ O-L ”.
<csum>
= checksum of string data. It is calculated by executing the exclusive OR (XOR) of all the characters
from STX (or from <Ind>) to ETX excluding the latter; the result of the XOR is decomposed into 2 characters
considering separately the 4 upper bits (first character) and the lower 4 bits (second character); the 2 charac-
ters obtained are then ASCII coded; (example: XOR = 5Dh; <csum> = "5Dh" that is 35h and 44h).