 12. Technical Data

74.4 mm (2.93 in.)

25.15 mm (.99 in.)

109.2 mm (4.3 in.)

26.4 mm (1.04 in.)

Dimensions

General

Polarization Source:

Signet 2650 Amperometric Electronics

Compatible Flow Cells

Signet 3-3610-1 (159 001 683)
Signet 3-3610-2 (159 001 684)
Signet 3-4630.392 (159 001 690)

Mounting

Signet DryLoc connection

Materials

CPVC

Free Chlorine and Chlorine Dioxide:

Membrane material:  PTFE

Free Chlorine and Chlorine Dioxide:

O-ring material:  FPM
Working electrode:  Gold
Counter reference electrode:  Silver halide

Wetted Materials:

PVC or PTFE, FPM, Nylon, Silicone

Performance

Electrode

:

Repeatability:  ± 0.08 ppm (mg/l) or 3% of selected 
range whichever is less
Free Chlorine Slope:  15 to 60 nA/ppm (mg/l) @ 25 ºC
Chlorine Dioxide Slope:  40 to 200 nA/ppm (mg/l) @ 17 ºC
Response time, T90:  < 2 minutes

System:

 (including electronics and instrument)

Accuracy:  < ±3% of electrode signal after calibration
Resolution:  

 0.5% of electrode range

Sensor Conditioning:

New, 

fi

 rst start-up:  4 hours maximum before calibration

Subsequent start-ups:  2 hours maximum

Temperature Element:

PT1000

Operational Ranges and Limits

Free Chlorine:

Range:  3-2630-1: 0.02 to 2 ppm (mg/L)
 

3-2630-2: 0.05 to 5 ppm (mg/L)

 

3-2630-3: 0.1 to 20 ppm (mg/L)

pH Operating Range:  5.5 pH to 8.5 pH

Chlorine Dioxide:

Range:  3-2632-1: 0.02 to 2 ppm (mg/L)
pH Operating Range:  4.0 pH to 11 pH

Maximum Operating Temperature:

0 °C to 45 °C (32 °F to 113 °F)

Maximum Operating Pressure:

Membrane:  0.48 bar @ 25 °C (7 psi @ 77 °F)

Flow Velocity Across Membrane Surface:

Minimum:  15 cm/s (0.49 ft/s)
Maximum:  30 cm/s (0.98 ft/s)

Free Chlorine Cross Sensitivity:

Chlorine Dioxide, ozone, bromine

Chlorine Dioxide Cross Sensitivity:

Free Chlorine, ozone

Chemical Compatibility:

< 50% ethanol/water, < 50% glycerol/water

Environmental Requirements

Storage Temperature (dry):

-10 ºC to 60 ºC (-4 ºF to 140 ºF)

System Temperature:

-10 ºC to 60 ºC (-4 ºF to 140 ºF)

Relative Humidity:

0 to 95% indoor/outdoor non-condensing to rated ambient

Standards and Approvals

CE, WEEE
Manufactured under ISO 9001
RoHS Compliant
China RoHS (Go to www.gfsignet.com for details)

 

Declaration of Conformity according to FCC Part 15

 

This device complies with Part 15 of the FCC rules.

 

Operation is subject to the following two conditions:

 

(1) This device may not cause harmful interference, and,

 

(2) This device must accept any interference received,

 

including interference that may cause undesired operation.

Summary of Contents for Signet 3-2630 Series

Page 1: ...vice time chlorine range high and low pH with optional Signet pH electrode temperature limits and more are stored on the chip This information is accessible via the Signet 8630 transmitter Signet s pa...

Page 2: ...rane caps carry no warranty Diluted HCl can irritate the eyes and skin use proper safety equipment Do not use surface tension reducing chemicals detergents or solvents on the membrane Electrode Range...

Page 3: ...greater than 2 months 4 If the sensor still fails to work properly continue with steps 11 17 Storage periods greater than 2 months If the sensor has been in storage for a long period of time or used...

Page 4: ...MERGE THE SENSOR See Figure 2 13 Apply power to the system 14 Monitor the nA of the sensor press the down arrow once on the 8630 transmitter The nA reading should start to rise Response time and nA re...

Page 5: ...The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 hypochlorous acid HOCl and hypochlorite OCl Above pH...

Page 6: ...n water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outside worki...

Page 7: ...Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 mL 3 2632 391 159 310 160 Chlorine Dioxide electrolyte solution 30 mL 3 2630 394 159 310 164 Free Chl...

Page 8: ...trode range Sensor Conditioning New first start up 4 hours maximum before calibration Subsequent start ups 2 hours maximum Temperature Element PT1000 Operational Ranges and Limits Free Chlorine Range...