background image

2

263X Series Chlorine Electrodes

Verifying the sensors accuracy using the DPD method should be performed to determine if the sensor requires maintenance.

1. 

Inspect the membrane for dirt or damage. Replace the membrane if its torn or if the gold cathode is visible. 

2. 

If the membrane is dirty clean the membrane by soaking it in 1 - 5% HCL and gently washed with a stream of DI water.

 

(do not use any mechanical device on the membrane)

 5.  Maintenance

Refi ll Procedure:

When adding electrolyte, be prepared for an accidental spill.
Working near a sink is recommended.

1.

  Remove the membrane cap from the front of the sensor.

2

.  Turn the sensor upside down and shake the sensor vigorously to remove the 

internal electrolyte.

3

.  Fill supplied syringe with electrolyte solution.

4.

  Place the electrode on a level surface.

5.

  Insert syringe needle fully into one of the eight electrode holes while injecting with 

electrolyte solution. Slowly injecting the electrolyte solution into the sensor to avoid 
introducing air bubbles. The electrode holds approximately 14 milliliters of solution. 
Slowly 

fi

 ll until solution begins to 

fl

 ow out of holes. Do not allow the solution to run 

down the electrode and wet the electrical contacts in the DryLoc connector.

6

.  Slowly screw on the membrane cap 

fi

 nger tight. Do not use tools.

To avoid damage and contamination, do not touch the white membrane surface 
on the membrane cap.

Avoid skin or eye contact with electrolyte solution.
Wear rubber gloves and goggles.

* Material Safety Data Sheets (MSDS) are available online at
www.gfsignet.com.

Additional caution should be taken when handling the 
Chlorine Dioxide electrolyte solution.

If a fresh water rinse does not clean the membrane, it will need to be replaced.
Keep spare membrane caps available. Membrane caps carry no warranty.

Diluted HCl can irritate the eyes and skin, use proper safety equipment.  
Do not use surface tension reducing chemicals, detergents or solvents on the membrane.

Electrode Range: 

The electrode must match the type and range of 

chlorine concentration to be measured.

Flow Rate:

The electrode must have a stable and constant 

fl

 ow of water past its 

membrane for accurate measurement. When the sensor is installed 
in the Tee 

fl

 ow cell 3-3610-2 (159 001 684), the 

fl

 ow rate must range 

from 37.8 to 75.7 LPH (10 to 20 US g/h).

When the sensor is installed in the Flow Cell Block 
3-4630.392 (159 001 690), the 

fl

 ow rate range should be 

30.24 to 45.36 LPH (8 to 12 US gph).

Sensor Conditioning: 4 hours

A new electrode requires conditioning of 4 hours with the electrode 
powered on and water 

fl

 owing past the head to generate a 

stable reading.

Subsequent start-ups can require an electrode conditioning of up to two hours.

 3.  Operation

The maximum allowable operating pressure 
must be less than 0.48 bar (7 psi). 
Higher pressures will damage the electrode.

The electrode should not be used in water 
containing surfactants, oils, organic chlorine 
or stabilizers such as cyanuric acid.

Part Number

Chlorine Range

Chlorine Type

3-2630-1

0.02 to 2 ppm (mg/L)

Free chlorine

3-2630-2

0.05 to 5 ppm (mg/L)

3-2630-3

0.1 to 20 ppm (mg/L)

3-2632-1

0.02 to 2 ppm (mg/L) Chlorine  Dioxide

 4.  Calibration

A new chlorine electrode or one that has had the membrane cap changed must be calibrated. Refer to the Signet 8630 Chlorine 
Transmitter manual for electrode and instrument calibration information. A diethyl-p-phenylenediamine (DPD) colorimeter test kit 
(not included) is required for sensor calibration. A sample is taken and analyzed with the DPD test kit, then this value is entered into 
the Signet 8630 transmitter.

•  Calibrate after a membrane cap change.
•  Check calibration one day after sensor is placed in service.
•  Check calibration weekly to monthly depending on process requirements.

Summary of Contents for Signet 3-2630 Series

Page 1: ...vice time chlorine range high and low pH with optional Signet pH electrode temperature limits and more are stored on the chip This information is accessible via the Signet 8630 transmitter Signet s pa...

Page 2: ...rane caps carry no warranty Diluted HCl can irritate the eyes and skin use proper safety equipment Do not use surface tension reducing chemicals detergents or solvents on the membrane Electrode Range...

Page 3: ...greater than 2 months 4 If the sensor still fails to work properly continue with steps 11 17 Storage periods greater than 2 months If the sensor has been in storage for a long period of time or used...

Page 4: ...MERGE THE SENSOR See Figure 2 13 Apply power to the system 14 Monitor the nA of the sensor press the down arrow once on the 8630 transmitter The nA reading should start to rise Response time and nA re...

Page 5: ...The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 hypochlorous acid HOCl and hypochlorite OCl Above pH...

Page 6: ...n water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outside worki...

Page 7: ...Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 mL 3 2632 391 159 310 160 Chlorine Dioxide electrolyte solution 30 mL 3 2630 394 159 310 164 Free Chl...

Page 8: ...trode range Sensor Conditioning New first start up 4 hours maximum before calibration Subsequent start ups 2 hours maximum Temperature Element PT1000 Operational Ranges and Limits Free Chlorine Range...

Reviews: