6
Signet PVDF Conductivity Sensors
To 8850-3
Sensr Gnd
(SHIELD)
Iso. Gnd
(BLACK)
Temp. IN
(WHITE)
Signal IN
(RED)
14
13
12
11
To 8860
SHLD
ISO GND
TEMP 1
SGNL 1
RED
WHITE
BLACK
SHIELD (Gnd)
SHLD
ISO GND
TEMP 1
SGNL 1
RED
WHITE
BLACK
SHIELD (Gnd)
Ch 1
Ch 2
• Do not route electrode cable in conduit containing AC power wiring. Electrical noise may interfere with electrode signal.
• Routing electrode cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• For resistivity measurements above 10 M
Ω
or below 20 °C, maximum cable length is 7.6 m (25 ft).
9900-1 Field Mount
with 8052 Integral Mounting Kit
and 9900.396 Angle Adjustment Adapter
Shield
T
emp. In
Iso. Ground
Signal
Black
Red
White
Shield
9900.394
Direct Conductivity/Resistivity Module
Input Connector
SW1
D3
1
2
3
1
2
3
4
ON
CTS
V+
DA
T
A
GND
SHLD
V+
Gnd
Data
Temp. IN
(WHITE)
Sensr Gnd
(SHIELD)
Signal IN
(RED)
Signal RTN
(BLK)
Black
Red
White
9900 S
3
L Input
via 2850 Sensor Electronics
General Wiring
ProcessPro Wiring
SmartPro Wiring