Signet PVDF Conductivity Sensors

To 8850-3

Sensr Gnd

(SHIELD)

Iso. Gnd

(BLACK)

Temp. IN

(WHITE)

Signal IN

(RED)

14

13

12

11

To 8860

SHLD

ISO GND

TEMP  1

SGNL 1

RED

WHITE

BLACK

SHIELD (Gnd)

SHLD

ISO GND

TEMP  1

SGNL 1

RED

WHITE

BLACK

SHIELD (Gnd)

Ch 1

Ch 2

•  Do not route electrode cable in conduit containing AC power wiring. Electrical noise may interfere with electrode signal.
•  Routing electrode cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
•  Seal cable entry points to prevent moisture damage.
•  For resistivity measurements above 10 M

Ω

 or below 20 °C, maximum cable length is 7.6 m (25 ft).

9900-1 Field Mount 

with 8052 Integral Mounting Kit

and 9900.396 Angle Adjustment Adapter

Shield

T

emp. In

Iso. Ground

Signal

Black

Red

White

Shield

9900.394

Direct Conductivity/Resistivity Module

Input Connector

SW1 

D3 

1

2

3

ON 

CTS 

V+

DA
T

A

GND

SHLD

V+

Gnd

Data

Temp. IN
(WHITE)

Sensr Gnd
(SHIELD)

Signal IN
(RED)

Signal RTN
(BLK)

Black

Red

White

9900 S

3

L Input

via 2850 Sensor Electronics

General Wiring

ProcessPro Wiring

SmartPro Wiring

Summary of Contents for Signet 2839

Page 1: ...cm 1 English Deutsch Fran ais Espa ol Italiano Warranty Information 2 Product Registration 2 Safety Information 2 Specifications 3 Dimensions 3 Cell Constant Selection 4 Custom Cell Certificate 4 Ins...

Page 2: ...d lead to failure of the retaining nut Do Not Use Tools Use of tool s may damage product beyond repair and potentially void product warranty Note Technical Notes Highlights additional information or d...

Page 3: ...0 s 2841 20 s 2842 30 s Operating Pressure Temperature With thread engagement per ANSI B1 20 1 10 C to 100 C 6 9 bar 14 F to 212 F 100 psi 10 C to 131 C 2 76 bar 14 F to 268 F 40 psi Storage Temperatu...

Page 4: ...be engaged to provide pressure capacity per published specifications The preferred installation for in line applications directs flow straight into the electrode This configuration dislodges entrappe...

Page 5: ...of the electrode fill the lower 75 mm to 100 mm 3 in to 4 in of conduit or extension pipe with a flexible sealant such as silicone 8850 3 Conductivity Transmitter locking ring conduit base 8052 integ...

Page 6: ...revent electrical noise and mechanical damage Seal cable entry points to prevent moisture damage For resistivity measurements above 10 M or below 20 C maximum cable length is 7 6 m 25 ft 9900 1 Field...

Page 7: ...before attempting maintenance Insulator removal and replacement A Thread the screw into the insulator 8 32 thread B Pull to remove the old insulator C Clean any coating or deposits inside the electrod...

Page 8: ...s 3 2850 51 40V 159 001 819 Integral 2850 system Digital S3 L output 0 1 cell PVDF NPT threads 3 2850 51 41V 159 001 820 Integral 2850 system Digital S3 L output 1 0 cell PVDF NPT threads 3 2850 51 42...

Reviews: