Signet PVDF Conductivity Sensors

Conductivity electrodes require little maintenance except for periodic cleaning in installations where contaminants are present.

•  Keep metal surfaces clean and free of coatings.

Replacement Insulator, 2842 electrode only

•  2842 electrodes have a removable insulator so the internal cavity can be cleaned.

•  After the insulator snaps into position it cannot be removed without damage.

•  Order insulator replacement kit 3-2842.390 before attempting maintenance.

Insulator removal and replacement

A.  Thread the screw into the insulator (8-32 thread).

B.  Pull to remove the old insulator.

C.  Clean any coating or deposits inside the electrode.

D.  Insert the new insulator and press into place.

Maintenance

2842 Insulator

NOTE:

Contact the manufacturer for 
replacement part number and cost.

9 mm

(0.34 in.)

41 mm (1.6 in.)

25.4 mm (1.0 in.)

14 mm

(0.55 in.)

3-2842.390 Insulator replacement kit

8-32 thread for removal

11 mm

(0.44 in.)

A

Electrode tip 

(side view)

Electrode

B

Insulator

C

Internal contact

Cotton swab

¥  Clean internal and external surfaces

Electrode tip

(end view)

D

New Insulator

electrode tip 

(side view)

Summary of Contents for Signet 2839

Page 1: ...cm 1 English Deutsch Fran ais Espa ol Italiano Warranty Information 2 Product Registration 2 Safety Information 2 Specifications 3 Dimensions 3 Cell Constant Selection 4 Custom Cell Certificate 4 Ins...

Page 2: ...d lead to failure of the retaining nut Do Not Use Tools Use of tool s may damage product beyond repair and potentially void product warranty Note Technical Notes Highlights additional information or d...

Page 3: ...0 s 2841 20 s 2842 30 s Operating Pressure Temperature With thread engagement per ANSI B1 20 1 10 C to 100 C 6 9 bar 14 F to 212 F 100 psi 10 C to 131 C 2 76 bar 14 F to 268 F 40 psi Storage Temperatu...

Page 4: ...be engaged to provide pressure capacity per published specifications The preferred installation for in line applications directs flow straight into the electrode This configuration dislodges entrappe...

Page 5: ...of the electrode fill the lower 75 mm to 100 mm 3 in to 4 in of conduit or extension pipe with a flexible sealant such as silicone 8850 3 Conductivity Transmitter locking ring conduit base 8052 integ...

Page 6: ...revent electrical noise and mechanical damage Seal cable entry points to prevent moisture damage For resistivity measurements above 10 M or below 20 C maximum cable length is 7 6 m 25 ft 9900 1 Field...

Page 7: ...before attempting maintenance Insulator removal and replacement A Thread the screw into the insulator 8 32 thread B Pull to remove the old insulator C Clean any coating or deposits inside the electrod...

Page 8: ...s 3 2850 51 40V 159 001 819 Integral 2850 system Digital S3 L output 0 1 cell PVDF NPT threads 3 2850 51 41V 159 001 820 Integral 2850 system Digital S3 L output 1 0 cell PVDF NPT threads 3 2850 51 42...

Reviews: