Signet PVDF Conductivity Sensors

•  Inspect threads to ensure integrity. Do not install an electrode with damaged threads.

•  Apply sealant or PTFE tape to threads.

•  Wetted materials include 316L stainless steel, PVDF and FKM (FKM O-ring inside 2841, 2842). Check 

for chemical compatibility before installing electrode.

•  Electrodes are supplied with 4.6 m (15 ft) of cable. It may be extended to a maximum 30 m (100 ft).

•  For resistivity measurements above 10 M

Ω

 or below 20 °C, maximum cable length is 7.6 m (25 ft).

•  When using the 2850 Conductivity/Resistivity Sensor Electronics, maximum cable length is 4.6 m (15 ft).

•  When using 9900, maximum cable length is 30 m (100 ft).

If the electrode is mounted vertically in a tee, do not recess the openings 
inside the tee. Mounting upside down may help prevent air entrapment.

An oversized tee may also be helpful for inline installations.

At least 4 threads (ANSI B1.20.1) must be engaged to provide pressure 
capacity per published specifi cations.

The preferred installation 

for in-line applications directs fl ow straight into the electrode. 

This confi guration dislodges entrapped air bubbles, and provides the best continuous sampling of the fl uid content.

The nominal process value should be near the center of the range.

Ranges below are for use with Signet Conductivity Instruments:

•  2839 (0.01 cm

-1

): 0.010 

to 

100 

μ

S (10 k

Ω

 to 100 M

Ω

)

•  2840 (0.1 cm

-1

): 1 

to 

1000 

μ

S (1 M

Ω

 to 1 k

Ω

)

•  2841 (1.0 cm

-1

): 10 

to 

10,000 

μ

S

•  2842 (10.0 cm

-1

): 100 

to 

200,000 

μ

S

(100 M

Ÿ

)

(10 k

Ÿ

)

2839 0.01 cm

-1

2841 1.0 cm

-1

2842 10.0 cm

-1

2840 0.1 cm

-1

0.010

μ

s 1

μ

s 10

μ

s 100

μ

s

1000

μ

s 10,000

μ

s

200,000

μ

s

Conductivity Range (

μ

S)

Signet Conductivity/Resistivity Electrode Ranges

Electrode Models

The sensor was delivered from the factory with a calibration certifi cate. 
The information provided on the certifi cate (custom cell constant and 
temperature offset) should be entered in the transmitter/controller. See 
individual product manual for details.

Cell Constant Selection

Custom Cell Certifi cate

Installation

In-Line Installation

Summary of Contents for Signet 2839

Page 1: ...cm 1 English Deutsch Fran ais Espa ol Italiano Warranty Information 2 Product Registration 2 Safety Information 2 Specifications 3 Dimensions 3 Cell Constant Selection 4 Custom Cell Certificate 4 Ins...

Page 2: ...d lead to failure of the retaining nut Do Not Use Tools Use of tool s may damage product beyond repair and potentially void product warranty Note Technical Notes Highlights additional information or d...

Page 3: ...0 s 2841 20 s 2842 30 s Operating Pressure Temperature With thread engagement per ANSI B1 20 1 10 C to 100 C 6 9 bar 14 F to 212 F 100 psi 10 C to 131 C 2 76 bar 14 F to 268 F 40 psi Storage Temperatu...

Page 4: ...be engaged to provide pressure capacity per published specifications The preferred installation for in line applications directs flow straight into the electrode This configuration dislodges entrappe...

Page 5: ...of the electrode fill the lower 75 mm to 100 mm 3 in to 4 in of conduit or extension pipe with a flexible sealant such as silicone 8850 3 Conductivity Transmitter locking ring conduit base 8052 integ...

Page 6: ...revent electrical noise and mechanical damage Seal cable entry points to prevent moisture damage For resistivity measurements above 10 M or below 20 C maximum cable length is 7 6 m 25 ft 9900 1 Field...

Page 7: ...before attempting maintenance Insulator removal and replacement A Thread the screw into the insulator 8 32 thread B Pull to remove the old insulator C Clean any coating or deposits inside the electrod...

Page 8: ...s 3 2850 51 40V 159 001 819 Integral 2850 system Digital S3 L output 0 1 cell PVDF NPT threads 3 2850 51 41V 159 001 820 Integral 2850 system Digital S3 L output 1 0 cell PVDF NPT threads 3 2850 51 42...

Reviews: