background image

3

Signet PVDF Conductivity Sensors

General

Instrument Compatibility

 .......2850, 8850-3, 8860, 9900

Cable

4.6 m (15 ft) std., 3 conductor with shield, 22 AWG, 
max length 30 m (100 ft). For resistivity measurements above 
10 M

Ω

 or below 20 °C, maximum cable length is 7.6 m (25 ft).

Process Connection

 -1V 

........................................

3

/

4

 in. NPT

 -1VD 

......................................ISO 

7/1-R 

3

/

4

Wetted materials

Threaded fi tting.........................PVDF
Insulator ....................................PVDF
Insulator O-ring (2841, 2842) ...FKM
Electrode contacts ....................316L stainless steel

Shipping Weight

2839..........................................0.34 kg (0.74 lb)
2840, 2841, 2842......................0.30 kg (0.66 lb)

Performance

Accuracy ...................................±2% of cell value
Temp. measurement .................PT1000
Temp. response time (T):
 2839 

......................................5 

s

 2840 

......................................10 

s

 2841 

......................................20 

s

 2842 

......................................30 

s

Operating Pressure / Temperature

With thread engagement per ANSI B1.20.1

  -10 ºC to 100 ºC @ 6.9 bar    (14 ºF to 212 ºF @ 100 psi)

  -10 ºC to 131 ºC @ 2.76 bar  (14 ºF to 268 ºF @ 40 psi)

Storage Temperature ................ -20 ºC to 131 ºC 

(-4 ºF to 268 ºF)

50

3

PSI

BAR

60

212

100

140

284

20

140

-20

-4

0

32 68

25

2

75

5

100

7

°F

°C

Standards and Approvals

  •     Manufactured under ISO 9001 for Quality, 

ISO 14001 for Environmental Management and 
OHSAS 18001 for Occupational Health and Safety.

  •   RoHS Compliant

 China RoHS (Go to www.gfsignet.com for details)

Specifi cations

59.3 mm
(2.33 in.)

73.7 mm
(2.90 in.)

140.7 mm

 (5.54 in.)

15.9 mm

(0.625 in.)

2839: 0.01 cm

-1

21.5 mm
(0.85 in.)

35.8 mm
(1.41 in.)

102.7 mm

 (4.05 in.)

15.9 mm

(0.625 in.)

2840: 0.1 cm

-1

15.9 mm

(0.625 in.)

26.8 mm
(1.13 in.)

109 mm

(4.29 in.)

41.9 mm
(1.65 in.)

2841: 1.0 cm

-1

2842: 10.0 cm

-1

Dimensions

Summary of Contents for Signet 2839

Page 1: ...cm 1 English Deutsch Fran ais Espa ol Italiano Warranty Information 2 Product Registration 2 Safety Information 2 Specifications 3 Dimensions 3 Cell Constant Selection 4 Custom Cell Certificate 4 Ins...

Page 2: ...d lead to failure of the retaining nut Do Not Use Tools Use of tool s may damage product beyond repair and potentially void product warranty Note Technical Notes Highlights additional information or d...

Page 3: ...0 s 2841 20 s 2842 30 s Operating Pressure Temperature With thread engagement per ANSI B1 20 1 10 C to 100 C 6 9 bar 14 F to 212 F 100 psi 10 C to 131 C 2 76 bar 14 F to 268 F 40 psi Storage Temperatu...

Page 4: ...be engaged to provide pressure capacity per published specifications The preferred installation for in line applications directs flow straight into the electrode This configuration dislodges entrappe...

Page 5: ...of the electrode fill the lower 75 mm to 100 mm 3 in to 4 in of conduit or extension pipe with a flexible sealant such as silicone 8850 3 Conductivity Transmitter locking ring conduit base 8052 integ...

Page 6: ...revent electrical noise and mechanical damage Seal cable entry points to prevent moisture damage For resistivity measurements above 10 M or below 20 C maximum cable length is 7 6 m 25 ft 9900 1 Field...

Page 7: ...before attempting maintenance Insulator removal and replacement A Thread the screw into the insulator 8 32 thread B Pull to remove the old insulator C Clean any coating or deposits inside the electrod...

Page 8: ...s 3 2850 51 40V 159 001 819 Integral 2850 system Digital S3 L output 0 1 cell PVDF NPT threads 3 2850 51 41V 159 001 820 Integral 2850 system Digital S3 L output 1 0 cell PVDF NPT threads 3 2850 51 42...