78812021-EN
CardioMem
®
CM 100 XT
45
13.9.1 Pre- and post-event times
The following settings are possible:
Pre/post (s)
25 / 15
45 / 15 *
30 / 30
60 / 30
60 / 60
Length of
ECG record-
ing (s)
40
60
60
90
120
(*) = default setting
The
“Length of ECG recording” results from the settings for pre-
and post-event times.
13.9.2 Settings for automatic detection of events
The following limit values and parameters can be set:
Parameter
Settable values
Default
Brady trigger
(bpm)
OFF, 30, 40, 50, 60
OFF
Tachy trigger
(bpm)
OFF, 100, 110, 120, 130, 140,
150, 160, 170, 180, 190, 200,
210, 220, 230, 240
OFF
Pause trigger
(seconds)
OFF; 2; 2.5; 3; 3.5; 4
OFF
AFib
OFF, ON
OFF
TIR
(time interval
recording)
OFF, 1 h ... 24 h (1h interval)
OFF
Tele ECG
OFF, ON
OFF