18 Troubleshooting
Symptom
Cause
Remedy
CM100 XT:
Device cannot be
turned on
(LED is not lit).
Low battery or no
battery inserted
Insert new battery.
CM 100 XT:
Recording can-
not be started
(LED is lit or-
ange).
Memory full
Connect the
recorder
to the
PC and the CM 100 Configu-
rator software and switch it
on.
Start CM 100 Configurator
and check the recorder sta-
tus.
Download the ECG data and
delete the memory of the re-
corder.
CM 100 XT:
Recording can-
not be started
(LED is flashing
orange, 3x beeps
when the
pushbutton is
pressed).
Open lead
Check the contacts of the
electrodes.
If an electrode became de-
tached, clean the skin and re-
place the electrode.
CM 100 XT:
LED is flashing
red.
Low battery
Replace the battery of the
CM 100 XT.
CM 100 XT:
Incomplete re-
cording
Low battery
Replace the battery of the
CM 100 XT.