Symptom
Cause
Remedy
(Error! No data
saved!).
CM 100
Configurator:
The ECG report
could not be cre-
ated.
(Error! No PDF
file created!).
Program installa-
tion defective or
damaged.
Install the CM 100 Configura-
tor anew. Repeat the down-
load.
CM 100
Configurator:
The ECG data
stored on the de-
vice cannot be
deleted.
(Error! Data not
deleted!).
Communication er-
ror
Click on
“Disconnect re-
corder
” and then on “Con-
nect to recorder
” to reset
the USB connection. If this
does not function:
Exit the CM 100 Configu-
rator software.
Remove the USB down-
load cable (the device is
turned off).
Reconnect the USB down-
load cable and restart the
recorder.
Start the CM 100 Configu-
rator software.
Click on
“Connect to re-
corder
” and repeat delet-
ing.
CM 100
Configurator:
The setup cannot
be transferred to
the device.
Communication er-
ror
Click on
“Disconnect re-
corder
” and then on “Connect
to recorder
” to reset the USB
connection. If this does not
function:
Exit the CM 100 Configu-
rator software.