13.4 Connecting the USB download cable to
recorder and PC
WARNING
RISK OF ELECTRIC SHOCK
– Only use the USB down-
load cable supplied by GETEMED to connect the re-
corder to a PC.
CAUTION
Make sure that the PC complies with the latest version of
the IEC 60950 standard for safety of IT equipment.
NOTICE
Connecting the recorder to a PC incorporated in an IT
network together with other equipment could result in pre-
viously unidentified risks to patients, users and third par-
ties. The responsible organisation must identify, analyse,
evaluate and control these risks.
Changes to the IT network, such as:
Changes in the configuration of the network/data
coupling configuration
Connection of additional components for net-
work/data coupling
Disconnection of components from network/data
coupling
Update of components connected to the net-
work/data coupling
Update of components connected to the net-
work/data coupling
could cause new risks that require additional analysis.
Refer to the standard EN 80001.
The download cable has two different connections - for the re-
corder and for the PC.
Insert the USB-A connector of the download cable into a free
USB port of your PC.