3.8 View-Only Menus
Three "view-only" menus (CAL, RELAY, and
OUTPUT) are available during normal operation.
Each view only menu provides the operator a
means of browsing through calibration settings
without dis-turbing system calibration and/or the
pH measurement process. When used in conjunc-
tion with the security feature, the view only menus
allow an op-erator to view most calibration infor-
mation on the front display, excluding the ability to
change sys-tem parameters without the supervisors
approval and personal security code.
Access each of the three view only menus by pres-
sing it's corresponding menu key. After entering
each of the three view only menus, each calibra-
tion parameter is sequentially displayed on the
main and lower displays by successively pressing
it's corresponding menu key as follows:
Main view-only menu:
pH sensor input option illustrated
Relay view-only menu:
2-Relay output option illustrated
Output view-only menu:
4 to 20 mA (iso/non-iso) output option illustrated
20
Note:
The view-only menus
are designed for viewing only
and DO NOT permit access
for calibration of any kind.
Menus will vary depending
upon installed options.
Exiting Menus:
Exit view
only menus at any time by
pressing:
pH
CAL
pH "Standard"
pH "Slope"
Temperature
mV input gain
setting
Relay 1
LO or HI setpoint
Relay 1
hysteresis
Relay 2
LO or HI setpoint
RELAY
Relay 2
hysteresis
OUTPUT
pH setting for
20 mA output
pH setting for
4 mA output