26
*Simulation signal polarity
MUST be observed. To
enable the negative polarity
indicator ( - ), shift the selected
display digit to the far left. To
disable (remove) the negative
polarity indicator, shift the
selected digit to the far left a
second time.
Voltage Input cards:
All voltge inputs MUST be
entered into memory as mV's
(i.e. 1.0 VDC = 1000 mV).
6. Enter the simulation signal value exactly as
displayed on the external digital multimeter
(DMM) as follows:
A. Press:
MOD
to enable 9030 display (display
will begin to flash)
B. Press:
mV
TEMP
pH
keys to enter the
exact*
simulation value. Press
ENTER
to confirm
entry.
Table 1
INPUT CARD:
INPUT TYPE:
4.0 pH SIMULATION:
10.0 pH SIMULATION:
(see Figure 16)
(step 5):
(step 8):
pH Sensor (mV)
A
3-2720 pH/ORP
Place pH sensor in
Place pH sensor in
Input Card
Preamplifier/Sensor
pH 4.0 buffer
pH 10.0 buffer
(#3-9030.400-1)
Record values:
B
mV Generator
Apply 177 mV
Apply -177 mV
4-20/0-20 mA
A
8710 pH/ORP
Place sensor
Place sensor
Input Card
Transmitter (unit must
in pH 4.0 buffer
in pH 10.0 buffer
(#3-9030.400-3)
be calibrated first,
see transmitter manual)
Jumper configurable
B
mA Generator
4-20 mA range:
4-20 mA range:
range, sec. 4.7
apply 8.6 mA
apply 15.5 mA
0-20 mA range:
0-20 mA range:
apply 5.7 mA
apply 14.3 mA
0-5/0-10 VDC
mV Generator
0-5 VDC range:
0-5 VDC range:
Input Card
apply 1400 mV
apply 3600 mV
(#3-9030.400-4)
(1.4 VDC)
(3.6 VDC)
Progammable range
0-10 VDC range:
0-10 VDC range:
sec. 3.2
apply 2300 mV
apply 7200 mV
(2.3 VDC)
(7.2 VDC)